Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review
Sanidhya Singh1, Miles Romney Bennett2, Chen Chen2
1University of Michigan Medical School, Ann Arbor, MI, United States.
Background:
Photoplethysmography (PPG) is a technology routinely used in clinical practice to assess blood oxygenation (SpO2) and pulse rate (PR). Skin pigmentation may influence accuracy, leading to health outcomes disparities.
Objective:
This systematic review and meta-analysis primarily aimed to evaluate the accuracy of PPG-derived SpO2 and PR by skin pigmentation. Secondarily, we aimed to evaluate statistical biases and the clinical relevance of PPG-derived SpO2 and PR according to skin pigmentation.
Methods:
We identified 23 pulse oximetry studies (n=59,684; 197,353 paired SpO2-arterial blood observations) and 4 wearable PR studies (n=176; 140,771 paired PPG-electrocardiography observations). We evaluated accuracy according to skin pigmentation group by comparing SpO2 accuracy root-mean-square values to the regulatory threshold of 3% and PR 95% limits of agreement values to +5 or -5 beats per minute (bpm), according to the standards of the American National Standards Institute, Association for the Advancement of Medical Instrumentation, and the International Electrotechnical Commission. We evaluated biases and clinical relevance using mean bias and 95% CI.
Results:
For SpO2, accuracy root-mean-square values were 3.96%, 4.71%, and 4.15%, and pooled mean biases were 0.70% (95% CI 0.17%-1.22%), 0.27% (95% CI -0.64% to 1.19%), and 1.27% (95% CI 0.58%-1.95%) for light, medium, and dark pigmentation, respectively. For PR, 95% limits of agreement values were from -16.02 to 13.54, from -18.62 to 16.84, and from -33.69 to 32.54, and pooled mean biases were -1.24 (95% CI -5.31 to 2.83) bpm, -0.89 (95% CI -3.70 to 1.93) bpm, and -0.57 (95% CI -9.44 to 8.29) bpm for light, medium, and dark pigmentation, respectively.
Conclusions:
SpO2 and PR measurements may be inaccurate across all skin pigmentation groups, breaching U.S. Food and Drug Administration guidance and industry standard thresholds. Pulse oximeters significantly overestimate SpO2 for both light and dark skin pigmentation, but this overestimation may not be clinically relevant. PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Pulse
The pulse serves as a clinical...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...