Related Experiment Video
Updated: Jun 10, 2025

A Model to Simulate Clinically Relevant Hypoxia in Humans
Published on: December 22, 2016
Oxygen reserve index vs. peripheral oxygen saturation for the prediction of hypoxemia in morbidly obese patients: a
Kemal Tolga Saraçoğlu1, Gülten Arslan2, Ayten Saraçoğlu3
1ICU and Perioperative Medicine Hazm Mebaireek General Hospital HMC, Industrial Area Ar- Rayyan Doha, Doha, Qatar. kemaltolgasaracoglu@gmail.com.
Background:
Pulse oximetry is a standart of anesthesia for perioperative monitoring. Due to the principles of Hb oxygen dissociation curve, peripheral oxygen saturation has an approximate sensitivity and specificity of 90% for the detection of hypoxemia.
Objectives:
The primary outcome of the study was to evaluate ORiⓇ as an early parameter to determine hypoxia in morbidly obese patients. The secondary outcome was to compare the effectiveness of ORiⓇ with SpO2 in non-obese patients.
Design:
Prospective, observational study.
Setting:
Department of elective operating room at tertiary hospital.
Patients And Methods:
Observational study included written informed consent from 51 patients with 19 < BMI < 25 kg/m2 and 51 patients with BMI > 40 kg/m2 undergoing an elective surgery requiring tracheal intubation. In addition to standard monitors, an ORi sensor was placed and baseline values were recorded. The patients were preoxygenated until end tidal expiratory oxygen concentration is reached to 90%. After anesthesia induction and tracheal intubation, the breathing circuit was not connected tracheal tube until the SpO2 decreased to 95%. Shapiro-Wilk, Pearson Chi-square, t-test, and Mann Whitney U test were used for the study.
Main Outcome Measures:
Times of tolerable apnea, ORiⓇ and SpO2 values at the end of preoxygenation, beginning of intubation, beginning of the ORi alarm, when SpO2 reached 95%, and when ORi reaches a plateau.
Sample Size:
102 patients.
Results:
The alert period: time to reach ORiⓇ from 0.24 to a value of 95% SpO2 was observed as 32 s in morbidly obese patients and 94 s in patients with a normal body mass index. The SpO2 alert period was determined as time difference between 97% and 95% SpO2. The data were recorded as 15 s and 36 s, respectively. It was observed that tolerable apnea, ORiⓇ, SpO2 and added alert times were longer in patients with normal BMI compared to morbidly obese patients.
Conclusions:
As a result, ORiⓇ can provide an early warning to prevent unexpected hypoxia before saturation begins to decrease in morbidly obese patients.
Limitations:
Inability to perform arterial blood gas sampling in the time periods when we looked at the parameters to determine the relationship between ORiⓇ and PaO2.
Gov Identifier:
NCT05480748 registered 2022-07-29.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
09:04Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
Related Concept Videos
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Oxygen Transport in the Blood