Validation and correlation of high-sensitive troponin I and troponin T in the emergency department
Moustafa Nseir1, Arash Mokhtari2,3, Mia Stanisic1
1Department of Internal and Emergency Medicine, Trelleborg Hospital, Trelleborg, Sweden.
Insights
High-sensitivity cardiac troponin T (hs-cTnT) elevations were twice as common as troponin I (hs-cTnI) in emergency department patients. This finding suggests hs-cTnT may have lower specificity and positive predictive value for acute myocardial infarction (AMI).
Area of Science:
- Cardiology
- Clinical Chemistry
- Emergency Medicine
Background:
- Troponin elevation is common in the Emergency Department (ED).
- Limited studies compare simultaneously measured high-sensitivity cardiac troponin T (hs-cTnT) and troponin I (hs-cTnI) in diverse clinical populations.
- Understanding differences in troponin assays is crucial for accurate diagnosis.
Purpose of the Study:
- To compare the performance of simultaneously measured hs-cTnT and hs-cTnI.
- To evaluate the diagnostic accuracy and prognostic implications of these assays in a broad patient cohort.
- To investigate the frequency of troponin elevations above the upper reference limit (URL).
Main Methods:
- Retrospective analysis of 1987 simultaneous hs-cTnT and hs-cTnI samples from 1134 patients.
- Inclusion of all patients undergoing troponin testing regardless of chief complaint.
- Statistical analysis of correlation, diagnostic accuracy (AUC, sensitivity, specificity, PPV, NPV), and 1-year mortality.
Main Results:
- A significant correlation was found between hs-cTnT and hs-cTnI (r=0.86, p<0.01).
- Elevated hs-cTnT (65%) was twice as frequent as elevated hs-cTnI (30%), with 39% showing elevated hs-cTnT and normal hs-cTnI.
- hs-cTnI demonstrated higher AUC (0.87) for AMI compared to hs-cTnT (0.80), with better specificity (80% vs 39%) and PPV (21.3% vs 9.3%).
- Both assays showed high negative predictive value (NPV) for AMI (98.5% for hs-cTnT, 98.3% for hs-cTnI).
- Elevated hs-cTnT and hs-cTnI were associated with increased 1-year mortality (HR 1.52 and 1.26, respectively).
Conclusions:
- Elevated troponin levels are common in a diverse ED population, especially hs-cTnT.
- hs-cTnT exhibits lower specificity and PPV for acute myocardial infarction (AMI) compared to hs-cTnI.
- The findings highlight potential differences in clinical utility between hs-cTnT and hs-cTnI assays for diagnosing AMI.
Background:
Troponin elevation is frequently observed in various scenarios in the Emergency Department (ED), yet there is a paucity of studies investigating simultaneously measured high-sensitivity cardiac troponin T (hs-cTnT) and troponin I (hs-cTnI) within a diverse cohort in a clinical setting.
Methods:
All patients who underwent troponin testing at a single center were eligible for this study. Only patients with simultaneous samples with hs-cTnI (Siemens) and hs-cTnT (Roche) were included, regardless of chief complaint.
Results:
Analysis of 1987 samples from 1134 patients showed a significant correlation between hs-cTnT and hs-cTnI (r = 0.86, p < 0.01). Of these samples, 65% exceeded the upper reference limit (URL) for hs-cTnT, and 30% for hs-cTnI with 39% who exhibited elevated hs-cTnT levels alongside normal hs-cTnI levels. The area under the curve (AUC) for acute myocardial infarction (AMI) for the index visit was 0.80 (95% CI; 0.75-0.85) for hs-cTnT and 0.87 (95% CI; 0.83-0.91) for hs-cTnI. Sensitivity and specificity were 91% and 39% for hs-cTnT, and 80% and 80% for hs-cTnI. Positive predictive value (PPV) and negative predictive value (NPV) was 9.3% and 98.5% for hs-cTnT respectively, corresponding for hs-cTnI was 21.3% and 98.3% respectively. Hazard ratios for 1-year mortality were 1.52 (95% CI; 1.40-1.66) for hs-cTnT and 1.26 (95% CI; 1.18-1.34) for hs-cTnI.
Conclusion:
Elevated troponins above the URL were very common in this diverse cohort, particularly for hs-cTnT, which was twice as frequent compared to hs-cTnI, resulting in low specificity and PPV for AMI.
More Related Videos
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
09:23Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic studies
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary SyndromeI: Introduction
Acute Coronary Syndrome II: Pathophysiology and clinical manifestations
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...