Related Experiment Video
Updated: Jun 10, 2025

Applications for Open Source Microplate-Compatible Illumination Panels
Published on: October 3, 2019
Evaluation of microplate handling accuracy for applying robotic arms in laboratory automation
Yoritaka Harazono1, Haruko Shimono2, Kikumi Hata2
1Department of Computational Biology and Medical Sciences, Graduate School of Frontier Sciences, University of Tokyo, 5-1-5, Kashiwanoha, Kashiwa, Chiba, 277-8561, Japan.
Abstract:
An inexpensive single-arm robot is widely utilized for recent laboratory automation solutions. The integration of a single-arm robot as a transfer system into a semi-automatic liquid dispenser without a transfer system can be realized as an inexpensive alternative to a fully automated liquid handling system. However, there has been no quantitative investigation of the positional accuracy of robot arms required to transfer microplates. In this study, we constructed a platform comprising aluminum frames and digital gauges to facilitate such measurements. We measured the position repeatability of a robot arm equipped with a custom-made finger by repeatedly transferring microplates. Further, the acceptable misalignment of plate transfer was evaluated by adding an artificial offset to the microplate position using this platform. The results of these experiments are expected to serve as benchmarks for the selection of robot arms for laboratory automation in biology. Furthermore, all information for replicating this device will be made publicly available, thereby allowing many researchers to collaborate and accumulate knowledge, hopefully contributing to advances in this field.