Related Experiment Video
Updated: Jun 10, 2025

Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
Visual and patient reported outcomes provided by a refractive multifocal intraocular lens based on continuous
Jorge L Alió1,2,3, Antonio Martínez-Abad4, Ramón Ruiz-Mesa5
1Vissum Miranza, Avda de Denia s/n Edificio Vissum, 03016, Alicante, Spain. jlalio@vissum.com.
Purpose:
To analyze the quality of vision of patients implanted bilaterally with the multifocal Precizon Presbyopic intraocular lens (IOL), as well as to evaluate the visual performance provided by the lens.
Setting:
Vissum Miranza Alicante.
Design:
Prospective multicenter study.
Methods:
56 patients (mean age 65.0 ± 8.7 years old) underwent bilateral implantation with multifocal Precizon Presbyopic IOL. The quality of vision was assessed by a quality of vision questionnaire at 6 months after the implantation procedure, a complete eye examination was also performed including visual and refractive measurements, defocus curve and contrast sensitivity assessment. Visual and refractive variables were compared in preoperative, 3-month postoperative and 6-month postoperative visits by Wilcoxon test.
Results:
The quality of vision analysis showed the absence of severe glare and severe haloes in all evaluated patients. Likewise, non-symptoms of glare, haloes and starbursts were seen in 75%, 68%, and 55% of subjects, respectively. Efficacy and safety index was 1.26 and 1.42, respectively. The 6-month postoperative binocular uncorrected distance visual acuity and near uncorrected visual acuity were 0.00 ± 0.09 and 0.20 ± 0.13 logMAR, respectively. Mean spherical equivalent was 0.29 ± 0.45 D.
Conclusions:
The Precizon Presbyopic NVA IOL (OPHTEC BV) provides a suitable quality of vision with a low rate of disturbance photic phenomena induction, as well as an excellent visual performance at main distances of sight accomplishing the visual demands of the majority of patients.