Introduction:
One-handedness is a challenge in conventional endoscopic ear surgery (EES). We present results on the first-ever application of the passive endoscope holder 'Endofix exo' (Co. AKTORmed GmbH, Neutraubling, Germany) in EES, which enables two-handed surgery.
Methods:
This two-sided study compares cut-suture time, operating time, postoperative complications, graft take rates, hearing results and quality of life in patients who underwent first stage tympanoplasty due to tympanic membrane perforation with intact ossicular chain conditions. 25 patients received classic EES (EES-, mean age: 28 ± 21 years) and 15 received EES with the passive holder (EES+, mean age: 48 ± 21 years).
Results:
Mean operating times (EES-: 96 ± 38 (SD) min; EES+: 107 ± 33 min), cut-suture times (EES-: 68 ± 30 min; EES+: 73 ± 31 min), complications, graft take rates and hearing results (preoperative air bone gap (ABG) (PTA4): 15 dB ± SD 8 dB (EES-); 16 dB ± SD 8 dB (EES+); postoperative ABG (PTA4): 11.25dB ± SD 11.3dB (EES-); 14 dB ± SD 10 dB (EES+)) did not differ significantly (p > 0.05) between the two groups. Postoperative hearing results and quality of life tended to improve in both groups (p > 0.05).
Discussion:
The passive endoscope holder has been successfully applied during the course of the study. However, modifications of the endoscope holder and further studies are recommended focusing on positioning of grafts and prostheses to obtain conclusive results regarding the superiority of two-handed EES over one-handed conventional EES.
Related Concept Videos
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...