Intermittent sigh breaths during high-frequency oscillatory ventilation in preterm infants: a randomised crossover
Judith Leigh Hough1,2,3, Luke Jardine4,5, Matthew James Hough6
1School of Allied Health, Australian Catholic University Faculty of Health Sciences, Banyo, Queensland, Australia judyhough@hotmail.com.
Insights
Adding sigh breaths to high-frequency oscillatory ventilation (HFOV) improved lung volume and oxygenation in preterm infants. This short-term intervention enhanced end-expiratory lung volume (EELV) and oxygen saturation in vulnerable newborns.
Area of Science:
- Neonatal Medicine
- Pediatric Respiratory Care
- Critical Care Medicine
Background:
- Preterm infants often require mechanical ventilation, with high-frequency oscillatory ventilation (HFOV) being a common mode.
- Maintaining adequate end-expiratory lung volume (EELV) and oxygenation is crucial for improving outcomes in these vulnerable patients.
- Conventional HFOV may not always ensure optimal lung recruitment and gas exchange.
Purpose of the Study:
- To investigate the efficacy of combining HFOV with intermittent sigh breaths.
- To assess the impact of this combined approach on EELV and oxygenation in preterm infants.
- To determine if sigh breaths enhance lung volume distribution during HFOV.
Main Methods:
- A prospective interventional crossover study was conducted in a neonatal intensive care unit.
- Ventilated preterm infants (<36 weeks corrected gestational age) receiving HFOV were randomly assigned to HFOV with sigh breaths or HFOV alone.
- Electrical impedance tomography (EIT) was used to measure EELV and ventilation distribution, with physiological variables recorded to assess oxygenation.
Main Results:
- The addition of sigh breaths to HFOV significantly increased global EELV (p=0.04).
- Ventilation distribution improved, with increased aeration in the posterior and left lung segments.
- Oxygen saturations showed a significant improvement (p<0.01) with the addition of sigh breaths.
Conclusions:
- Intermittent sigh breaths during HFOV are associated with short-term improvements in EELV.
- This strategy enhances lung aeration in dependent and left lung regions.
- The combination of HFOV and sigh breaths leads to improved oxygenation in preterm infants.
Objective:
To determine if combining high-frequency oscillatory ventilation (HFOV) with additional sigh breaths would improve end-expiratory lung volume (EELV) and oxygenation in preterm infants.
Design:
Prospective interventional crossover study.
Setting:
Neonatal intensive care unit.
Patients:
Ventilated preterm infants <36 weeks corrected gestational age receiving HFOV.
Interventions:
Infants were randomly assigned to receive HFOV with sigh breaths followed by HFOV-only (or vice versa) for four alternating periods. Sigh breaths were delivered with an inspiratory time of 1 s, peak inspiratory pressure of 30 cmH20 and frequency of three breaths/min.
Main Outcome Measures:
Electrical impedance tomography measured the effect of sigh breaths on EELV and ventilation distribution. Physiological variables were recorded to monitor oxygenation. Measurements were taken at 30 and 60 min postchange of HFOV mode and compared with baseline.
Results:
Sixteen infants (10 males, 6 females) with a median (range) gestational age at birth of 25.5 weeks (23-31), study weight of 950 g (660-1920) and a postnatal age of 25 days (3-49) were included in the study. The addition of sigh breaths resulted in a significantly higher global EELV (mean difference±95% CI) (0.06±0.05; p=0.04), with increased ventilation occurring in the posterior (dependent) and left lung segments, and improved oxygen saturations (3.31±2.10; p<0.01).
Conclusion:
Intermittent sigh breaths during HFOV were associated in the short-term with an increased EELV in the posterior and left lungs, and improved oxygen saturations in preterm infants.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...
Pulmonary Cycle: Exhalation
Respiratory Volumes and Capacities I
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...