Related Experiment Video
Updated: Jun 30, 2026

A Postoperative Evaluation Guideline for Computer-Assisted Reconstruction of the Mandible
Published on: January 28, 2020
Passive fit and time efficiency for prefabricated versus conventionally constructed cobalt chromium CAD\CAM 3-unit
Mohamed El-Sayed Kamel1, AlHassan Alaa Eldin Alsayed1, Mohamed Amr ElKhashab2
1Master Degree Candidate of Oral Implantology, Faculty of Dentistry, Cairo Univeristy, Cairo, Egypt.
Background:
The passive fit of 3-unit implant supported prefabricated metal screw-retained prosthesis before implant placement might be difficult. Hence, we aim to evaluate the passive fit and time efficiency of CAD/CAM 3-unit implant supported fixed prostheses that were constructed based on virtual versus those based on actual implant positions in Kennedy Class I models.
Methods:
A sample of 5 Kennedy class I models with thin wiry ridges were restored by 20 frameworks bilaterally, 10 based on actual (group A) and 10 based on virtual (group V) implant positions. The models were imaged using cone beam computed tomography and scanned using an intraoral scanner. The STL (Standard Tessellation Language files) and the DICOM (Digital Imaging and Communications in Medicine) files were registered on a 3D planning software. A CAD/CAM surgical guide was planned, resin printed and used for installing 6 implants bilaterally. In group V, the framework was designed based on the virtual scan bodies and virtual multi-unit abutments, while in group A intra-oral scanning of the model after attaching the scan bodies was necessary. Frameworks of both groups were milled and tested for passive fit using 8 clinical tests. McNemar and Wilcoxon signed rank tests were used to study the effect of the group on passive fit and time efficiency, respectively. The significance level was set at P ≤ 0.05.
Results:
No statistically significant difference was found between group V and group A frameworks regarding passive fit (p-value = 1, OR = 0.5) and time efficiency (P = 0.179, Effect size = 0.948).
Conclusion:
Within the limitations of this study, it can be concluded that in free end saddle cases, prefabricated CAD\CAM 3-unit implant-supported cobalt chromium screw retained prostheses can achieve an adequate passive fit. However, their fit might be negatively affected in thin ridges and they might require some adjustments.
Related Concept Videos
Support Reactions
The purpose of the supports is to prevent the translational motion of the system by applying an equal and opposite force and to prevent the system's rotation by applying a...
Frames
Frames are versatile and widely used in various applications such as structural supports for beams and columns, automobile chassis construction, and in the construction...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Design of Prismatic Beams for Bending