Elisabet Navarro-Tapia1,2, Ana Herranz Barbero3, Maribel Marquina4
1Faculty of Health Sciences, Valencian International University (VIU), Valencia, Spain.
Frontiers in Pediatrics
|October 16, 2024
Summary
Mid-infrared analysis (MIRIS-HMA) accurately measures human milk protein concentration, offering a reliable alternative to traditional methods for neonatal nutrition. This method shows better agreement with the Bradford assay for fortifying donor milk.
Area of Science:
- Neonatal Nutrition
- Analytical Chemistry
- Biotechnology
Background:
- Human milk is optimal for newborns, especially premature infants.
- Donor human milk is a viable alternative when mother's milk is unavailable.
- Nutritional fortification of human milk is often necessary to meet infant needs.
Purpose of the Study:
- To evaluate the concordance of two novel analyzers (mid-infrared and ultrasound) with the Bradford method for human milk protein quantification.
- To assess the feasibility of implementing faster, simpler analyzers in neonatal healthcare settings.
- To determine the accuracy of MIRIS-HMA and ultrasound methods for protein analysis in donor human milk.
Main Methods:
- Protein concentration in 142 donor human milk samples was measured using mid-infrared (MIRIS-HMA), ultrasound (MilkoScope), and Bradford assays.
- Intraclass correlation coefficient (ICC) and Bland-Altman plots were employed to assess method agreement.
- Milk samples were collected and pasteurized at a human milk bank.
Main Results:
- Mean protein concentrations were 1.38 g/100ml (MIRIS-HMA), 1.15 g/100ml (MilkoScope), and 1.19 g/100ml (Bradford).
- MIRIS-HMA demonstrated a higher ICC (0.70) compared to MilkoScope (0.37) when correlated with the Bradford method.
- MIRIS-HMA showed superior agreement with the reference Bradford assay.
Conclusions:
- MIRIS-HMA is a promising method for accurate human milk protein analysis, aligning well with the Bradford technique.
- The findings support the development of new devices based on mid-infrared transmission spectroscopy for neonatal applications.
- MIRIS-HMA offers a reliable approach for protein quantification, crucial for optimizing fortified human milk for vulnerable infants.