Optimization of image reconstruction technique for respiratory-gated lung stereotactic body radiotherapy treatment
Kenji Yasue1, Hiraku Fuse2, Minori Takaoka3
1Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, 4669-2, Ami, Inashiki, Ibaraki, 300-0394, Japan. yasueke@ipu.ac.jp.
Abstract:
Patient respiration is characterized by respiratory parameters, such as cycle, amplitude, and baseline drift. In treatment planning using four-dimensional computed tomography (4DCT) images, the target dose may be affected by variations in image reconstruction techniques and respiratory parameters. This study aimed to optimize 4DCT image reconstruction techniques for the treatment planning of lung stereotactic body radiotherapy (SBRT) based on respiratory parameters using respiratory motion phantom. We quantified respiratory parameters using 30 respiratory motion datasets. The 4DCT images were acquired, and the phase- and amplitude-based reconstruction images (RI) were created. The target dose was calculated based on these reconstructed images. Statistical analysis was performed using Pearson's correlation coefficient (r) to determine the relationship between respiratory parameters and target dose in each reconstructed technique and respiratory region. In the inhalation region of phase-based RI, r of the target dose and baseline drift was -0.52. In particular, the target dose was significantly reduced for respiratory parameters with a baseline drift of 0.8 mm/s and above. No other respiratory parameters or respiratory regions were significantly correlated with target dose in phase-based RI. In amplitude-based RI, there were no significant differences in the correlation between all respiratory parameters and target dose in the exhalation or inhalation regions. These results showed that the target dose of the amplitude-based RI did not depend on changes in respiratory parameters or respiratory regions, compared to the phase-based RI. However, it is possible to guarantee the target dose by considering respiratory parameters during the inhalation region of the phase-based RI.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography