From age-specific to size-specific dose protocol for paediatric head computed tomography: a simple practical strategy
Liwei Hu1,2, Rongzhen Ouyang1, Jing Li2,3
1Department of Radiology, Shanghai Children's Medical Center, School of Medicine, Shanghai Jiao Tong University, 1678 Dongfang Road, Shanghai 200127, China.
This study introduces a method to transition pediatric head CT scans from age-based to size-based radiation dose protocols. Findings suggest certain pediatric groups may receive unnecessary radiation exposure, necessitating protocol adjustments.
Area of Science:
- Medical Imaging
- Radiology
- Pediatric Health
Background:
- Current pediatric head CT protocols often rely on age-specific radiation doses.
- Variations in patient size within age groups can lead to suboptimal radiation exposure.
- A need exists for optimized dose protocols in pediatric computed tomography (CT).
Purpose of the Study:
- To develop a practical approach for transitioning from age-specific to size-specific dose protocols in pediatric head CT.
- To identify appropriate transitioning parameters for optimizing radiation dose in pediatric patients.
- To evaluate the necessity and effectiveness of size-specific dose protocols.
Main Methods:
- Utilized two existing age-specific protocols for pediatric head CT (1-6 years and >6 years).
- Defined a characteristic head size threshold (HDthreshold) based on anteroposterior head diameter.
- Categorized 135 pediatric cases into four groups based on age and head size for analysis.
- Assessed image quality using both subjective and objective methods.
Main Results:
- Identified potential radiation overdose in Group C (patients >6 years, ≤HDthreshold) by 38%.
- Suggested that the exposure settings for Group D (patients >6 years, >HDthreshold) could replace those used for Group B (patients ≤6 years, >HDthreshold).
- Demonstrated the sufficiency of the proposed strategy in assessing the need for and determining parameters for size-specific dose protocols.
Conclusions:
- The proposed strategy effectively assesses the need for transitioning to size-specific dose protocols in pediatric head CT.
- Determining transitioning parameters based on head size alongside age can optimize radiation exposure.
- Implementation of size-specific protocols can help mitigate potential radiation overdose in pediatric patients.