Ultra-Low-Field Portable MRI and Extracorporeal Membrane Oxygenation: Preclinical Safety Testing
Jin Kook Kang1, Eric Etchill1, Kate Verdi1
1Division of Cardiac Surgery, Department of Surgery, Johns Hopkins Hospital, Baltimore, MD.
Context:
Conventional MRI is incompatible with extracorporeal membrane oxygenation (ECMO) cannulas and pumps. Ultra-low-field portable MRI (ULF-pMRI) with 0.064 Tesla may provide a solution, but its safety and compatibility is unknown.
Hypothesis:
ULF-pMRI does not cause significant displacement and heating of ECMO cannulas and does not affect ECMO pump function.
Methods And Models:
ECMO cannulas in various sizes were tested ex vivo using phantom models to assess displacement force and heating according to the American Society for Testing and Materials criteria. ECMO pump function was assessed by pump flow and power consumption. In vivo studies involved five female domestic pigs (20-42 kg) undergoing different ECMO configurations (peripheral and central cannulation) and types of cannulas with an imaging protocol consisting of T2-weighted, T1-weighted, FLuid-Attenuated Inversion Recovery, and diffusion-weighted imaging sequences.
Results:
Phantom models demonstrated that ECMO cannulas, both single lumen with various sizes (15-24-Fr) and double lumen cannula, had average displacement force less than gravitational force within 5 gauss safety line of ULF-pMRI and temperature changes less than 1°C over 15 minutes of scanning and ECMO pump maintained stable flow and power consumption immediately outside of the 5 gauss line. All pig models showed no visible motion due to displacement force or heating of the cannulas. ECMO flow and the animals' hemodynamic status maintained stability, with no changes greater than 10%, respectively.
Interpretation And Conclusions:
ULF-pMRI is safe and feasible for use with standard ECMO configurations, supporting its clinical application as a neuroimaging modality in ECMO patients.