Feasibility of ultra-low-activity 18F-FDG PET/CT imaging in children
Ha Wu1,2,3,4,5, Guobing Liu1,2,3,4, Haojun Yu1,2,3,4
1Department of Nuclear Medicine, Zhongshan Hospital, Fudan University, Shanghai, 200032, China.
Objectives:
To investigate the feasibility of paediatric 18F-FDG total-body PET/CT imaging with an ultra-low activity and explore an optimized acquisition time range.
Methods:
A total of 38 paediatric patients were prospectively enrolled and underwent dynamic total-body PET/CT imaging using ultra-low 18F-FDG activity (0.37 MBq/kg). The 60-minute list-mode raw data were acquired and then reconstructed as static PET images by using 50-51, 50-52, 50-53, 50-54, 50-55, 50-58, 50-60, and 45-60 minutes data, which were noted as G1, G2, G3, G4, G5, G8, G10, and G15, respectively. Image qualities were subjectively evaluated using the Likert scale and were objectively evaluated by the quantitative metrics including standard uptake value (SUV), signal-to-noise ratio (SNR), target-to-background ratio (TBR), and contrast-to-noise ratio (CNR).
Results:
The injected activity of FDG was 13.38 ± 5.68 MBq (4.40-28.16 MBq) and produced 0.58 ± 0.19 mSv (0.29-1.04 mSv) of effective dose. The inter-reader agreement of subjective image quality was excellent (kappa = 0.878; 95% CI, 0.845-0.910). The average scores of image quality for G1-G15 were 1.10 ± 0.20, 2.03 ± 0.26, 2.66 ± 0.35, 3.00 ± 0.27, 3.32 ± 0.34, 4.25 ± 0.30, 4.49 ± 0.36, and 4.70 ± 0.37, respectively. All image scores are above 3, and all lesions are detectable starting from G8. SNRs of backgrounds, TBRs, and CNRs were significant differences from the control group before G8 (all P < 0.05).
Conclusion:
The image quality of the 8 min acquisition for paediatric 18F-FDG total-body PET/CT with an ultra-low activity could meet the diagnostic requirements.
Advances In Knowledge:
This study confirms the feasibility of ultra-low dose PET imaging in children, and its methods and findings may guide clinical practice. Paediatric patients will benefit from reduced radiation doses.
More Related Videos
10:53Functional Imaging of Brown Fat in Mice with 18F-FDG micro-PET/CT
Published on: November 23, 2012
10:02Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography FDG-PET/CT
Published on: May 2, 2012
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET