The use of a novel toe-thumb pressure index for assessing arterial status in the lower limb. A reliability and
Juliana Mazzeo1, Helen A Banwell1, Peta E Tehan2,3
1Allied Health and Human Performance, Innovation, Implementation and Clinical Translation in Health (IIMPACT), University of South Australia, Adelaide, South Australia, Australia.
Aims:
This study explored the reliability, validity and perceived comfort of a novel thumb pressure measure and calculation of a toe-thumb index to identify their suitability as an adjunct or alternatives to ankle-brachial and toe-brachial indices.
Methods And Results:
Repeated manual thumb and toe systolic blood pressures were conducted using two raters, over two time points, on 34 healthy participants. Concurrent automated toe, thumb and brachial systolic blood pressures as well as comfort ratings for these measures (using a 10 mm visual analogue scale) were captured once by a research assistant. Automated thumb and brachial measures showed fair correlation (ρ = 0.36, p = 0.03) and a toe-thumb index and toe-brachial index good correlation (ρ = 0.62, p < 0.01). Intraclass correlation coefficients (ICC) identified moderate intra-rater reliability for manual thumb pressures for Rater 1 and 2 (ICC 0.57, 95% CI [0.14, 0.79] and ICC 0.74, 95% CI [0.49, 0.87], respectively), while inter-rater reliability was poor (ICC = 0.16, 95% CI [-0.85, 0.47]). Concurrent validity comparing manual and automated measures for thumb pressure was also poor (ICC -0.05, 95% CI [-1.06, 0.72] and ICC 0.42, 95% CI [-0.16, 0.72] Rater 1 and 2 respectively). Thumb measures were significantly more comfortable than brachial measures (5 mm, p < 0.00).
Conclusion:
Thumb systolic pressures are correlated with brachial systolic pressures, with reasonable intra-rater reliability, however, correlation is only fair and measurement error wider than clinically acceptable. Furthermore, manual measures are poorly correlated with automated units. Consequently, caution is required in applying these techniques. As thumb measures were perceived as significantly more comfortable than brachial measures and have an advantage where brachial pressures cannot, or should not, be obtained, further evaluation is warranted.
Related Concept Videos
Pulse Assessment Sites
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Assessing Blood pressure in the Leg
Preparation:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Pre-Procedural Guidelines for Assessing Blood Pressure