Rapid Antimicrobial Susceptibility Testing by Stimulated Raman Scattering Imaging of Deuterium Incorporation in a Single Bacterium
Published on: February 14, 2022
Detection of carbapenem-resistant gram-negative bacilli in Japan using the fully automated bacterial testing device
Yumiko Funashima1, Rin Hamabe2, Kei Tominaga2
1Department of Medical Technology and Science, Faculty of Fukuoka Health Care, International University of Health and Welfare, Okawa, Fukuoka, Japan; Medical Laboratory Science Graduate School of Health and Welfare Sciences, International University of Health and Welfare Graduate School, Okawa, Fukuoka, Japan.
Objective:
We investigated a rapid detection method for carbapenemase-producing gram-negative bacilli (CP-GNR) using meropenem (MEPM) to assess the efficiency of the antimicrobial susceptibility testing.
Methods:
We used the function that can monitor the growth curve with the resistant bacteria monitoring function (RAISUS S4). Rapid detection of CP-GNR was performed using RAISUS S4 in two types of antimicrobial susceptibility testing, the RAISUS 18-hour method (18-h method) and RAISUS rapid method (rapid method) for Enterobacterales (F-GNR) and non-fermenting gram-negative bacilli (NF-GNR).
Results:
When F-GNR were based on MEPM MIC ≥ 0.25 µg/mL, CP-GNR were detected with a sensitivity of 100% (58/58) for the 18-h method and 98.3% (57/58) for the rapid method; the shortest detection times were 5.3 and 4.0 h, respectively. When NF-GNR were based on MEPM MIC > 8 µg/mL, it was possible to detect CP-GNR with 100% sensitivity (58/58) in both methods. Furthermore, in the analysis using the 18-h method for monitoring resistant bacteria, when ≥ 2 µg/mL was used as the screening concentration for F-GNR, approximately 50% of the resistant genotypes, NDM, GES, and KPC, were detected in approximately 7 h However, detecting the IMP and VIM took 11-12 h CONCLUSIONS: The 18-h and rapid methods with RAISUS S4 were highly correlated with the results of the microdilution method of CLSI, and CP-GNR detection was rapid using a function that can monitor the growth curve with RAISUS S4.