Background:
Atrioventricular block is common with adult congenital heart disease and pacemaker implantation is challenging. Atrioventricular synchronous pacing is important for better haemodynamics. This case reports the implantation of a dual-chamber leadless pacemaker in a patient with univentricular heart physiology and contributes to the literature regarding the management option in complex adult congenital heart disease patients with conduction abnormalities.
Case Summary:
A 25-year-old male with double inlet left ventricular, transposition of great arteries, hypoplastic aortic arch receive multiple surgeries including the Glenn shunt at the age of 1. He presented with 2:1 and 3:1 heart block at the age of 13 with a transvenous dual-chamber pacemaker implanted by pacing the superior vena cava stump and puncturing the Glenn shunt for the ventricular lead. A decade later, lead malfunctioned and the patient progressed to complete heart block. A subcutaneous implantable cardioverter defibrillator was implanted when he was 23 for monomorphic ventricular tachycardia. Given the anticipated challenges with transvenous lead extraction and epicardial pacemaker implantation, we implanted the novel dual-chamber leadless pacemakers which resulted in satisfactory atrioventricular synchronous pacing performance immediately post-op and 2 weeks after the procedure.
Discussion:
We present a case of a novel dual-chamber leadless pacemaker implantation to maintain atrioventricular synchrony in the patient with complete heart block and univentricular physiology. This case illustrates an additional pacing option in complex adult congenital heart to maintain atrioventricular synchrony.