Development of a Polymersome Blood Ammonia Assay Coupled with a Portable Near-Infrared Fluorometer
Marie-Lynn Al-Hawat1, Justine Caron1, Sarah Djebbar1
1Faculté de Pharmacie, Université de Montréal, Montreal, Quebec H3T 1J4, Canada.
Abstract:
Ammonia is a key biomarker in inborn and acquired liver disease. As clinical point-of-care blood ammonia assays are lacking, we developed a polymersome formulation for point-of-care blood ammonia sensing combined with a portable fluorometer. A pH-sensitive near-infrared (NIR) fluorescent dye was identified, which showed a strong fluorescence increase at acidic pH values. Building on reports on ammonia-selective PS-b-PEG polymersomes, these polymersomes were loaded with the NIR dye. These NIR fluorescent polymersomes sensed ammonia in a clinically relevant range in ammonia-spiked fresh whole blood with high linearity (R 2 = 0.9948) after 5 min using a conventional tabletop plate reader. Subsequently, the assay was tested with a portable fluorometer. An ammonia-dependent fluorescence increase was detected in ammonia-spiked fresh mouse blood after 5 min using the portable fluorometer. The NIR dye-loaded PS-b-PEG polymersomes rapidly sensed ammonia with high linearity in whole blood. This assay was successfully combined with a portable fluorometer and only required 3 μL of blood. These findings motivate a further development and clinical translation of this point-of-care blood ammonia assay.