Occupational Radiation Exposure During Intraoperative 3-Dimensional Fluoroscopy in Pelvis and Acetabular Surgery.
Michelle M Gosselin1, Daniel H Ngo2, Ashley E Levack3
1Department of Orthopaedics, Naval Medical Center San Diego, San Diego, CA.
Journal of Orthopaedic Trauma
|October 21, 2024
Summary
Three-dimensional (3D) fluoroscopy in pelvic and acetabular surgery shows occupational radiation exposure within safe limits. The study confirms that the radiation dose from 3D fluoroscopic spins is acceptable for operating room personnel.
Area of Science:
- Orthopedic Surgery
- Medical Imaging
- Radiation Safety
Background:
- Intraoperative imaging is crucial for complex orthopedic procedures.
- Three-dimensional (3D) fluoroscopy offers enhanced visualization but raises concerns about radiation exposure.
- Pelvic and acetabular surgeries often require extensive intraoperative imaging.
Purpose of the Study:
- To quantify occupational radiation exposure during 3D fluoroscopy in pelvic and acetabular surgery.
- To assess if the radiation dose from the 3D fluoroscopic spin is within established safety limits.
- To evaluate the safety of utilizing 3D fluoroscopy in orthopedic operating rooms.
Main Methods:
- Prospective cohort study conducted at a Level I trauma center.
- Inclusion of all patients undergoing 3D fluoroscopy for pelvic or acetabular surgery.
- Measurement of radiation dose at standardized locations during the 3D fluoroscopic spin phase.
Main Results:
- Seventy-six 3D spins were performed on 69 patients.
- The average displayed radiation dose during the spin was 39.9 mGy, 53.6% less than routine fluoroscopy.
- Average patient exposure was 3.42 mGy, with maximal room exposure at 0.062 mGy; minimal radiation detected outside the OR.
Conclusions:
- Occupational radiation exposure from 3D fluoroscopy in these surgeries is within safe limits.
- The radiation increase associated with 3D imaging is marginal and acceptable.
- The use of intraoperative 3D fluoroscopy in pelvic and acetabular surgery is encouraged.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
149
The most common cardiovascular diagnostic test is an X-ray. It produces images of the heart, blood vessels, and adjacent structures.
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
149
Radiological Investigation III: Pulmonary Angiogram and PET Scan
86
Radiological investigations are paramount in the diagnosis and management of various pulmonary diseases. Two essential investigations are the Pulmonary Angiogram and the Positron Emission Tomography (PET) Scan.
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
86