Related Experiment Video
Updated: Jun 9, 2025

Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Occupational Radiation Exposure During Intraoperative 3-Dimensional Fluoroscopy in Pelvis and Acetabular Surgery
Michelle M Gosselin1, Daniel H Ngo2, Ashley E Levack3
1Department of Orthopaedics, Naval Medical Center San Diego, San Diego, CA.
Objectives:
To quantify the occupational radiation exposure with a 3-dimensional (3D) fluoroscopic machine during routine use in pelvic and acetabular surgery and to determine whether the additional radiation exposure encountered with the 3D fluoroscopic spin is within previously accepted limits.
Design:
Prospective cohort study.
Setting:
Level I trauma center.
Patient Selection Criteria:
All patients undergoing 3D fluoroscopy intraoperatively during pelvis (OTA/AO 61B,C) or acetabular (OTA/AO 62A-C) surgery between April 2021 and July 2021.
Outcome Measures And Comparisons:
Radiation dose at standardized locations around the operating room during the spin portion of the 3D fluoroscopy.
Results:
Seventy-six 3D spins were performed on 69 patients during the study period. The average emitted radiation dose from the machine for the routine fluoroscopy portion of the case was 74.5 mGy. The average displayed radiation dose in the air for the spin portion of the case was 39.9 mGy, an average of 53.6% less radiation than the routine fluoroscopy portion. For the spin portion, the average radiation exposure seen by the patient was 3.42 mGy (centered on the patient) and the average maximal exposure in the room was 0.062 mGy. Minimal radiation was detected outside the operating room doors.
Conclusions:
The radiation exposure encountered by operating room personnel with 3D fluoroscopy appears to be within safe occupational limits. The marginal increase in radiation exposure during pelvic and acetabular surgery should not discourage the use of 3D imaging intraoperatively.
Level Of Evidence:
Level IV, Case Series.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...