Extracellular Wire Tetrode Recording in Brain of Freely Walking Insects
Published on: April 1, 2014
16.6K
Removal of errant coils using the Tigertriever-13
Shah Islam1, Ze'ev Itzakzonhayosh2, Emily Chung2
1Joint Department of Medical Imaging, Toronto Western Hospital Division of Neuroradiology, Toronto, Ontario, Canada s.islam@imperial.ac.uk.
Journal of Neurointerventional Surgery
|October 21, 2024
Summary
A new technique using the Tigertriever-13 device effectively retrieves migrated coils during neurovascular embolization. This low-profile method avoids the need for larger catheters in emergency situations, improving patient safety.
Area of Science:
- Interventional Neuroradiology
- Endovascular Surgery
- Medical Device Technology
Background:
- Coil migration is a recognized risk during embolization procedures.
- Existing retrieval methods, like stent retrievers, may be too large for small catheters.
- This limitation can necessitate emergency catheter upsizing.
Purpose of the Study:
- To describe and demonstrate a novel technique for retrieving migrated coils.
- To showcase the utility of the low-profile Tigertriever-13 device for coil retrieval.
- To present an in vivo application of the technique as a bailout strategy.
Main Methods:
- Ex vivo demonstration using silicone flow models.
- In vivo application during acute embolization of a ruptured peri-callosal aneurysm.
- Utilized the Tigertriever-13 device for retrieval of embolization coils.
Main Results:
- Successful retrieval of migrated coils was achieved using the Tigertriever-13 device.
- The device demonstrated feasibility in both ex vivo and in vivo settings.
- The technique served as an effective bailout during a complex aneurysm embolization.
Conclusions:
- The Tigertriever-13 device offers a viable low-profile solution for migrated coil retrieval.
- This technique can be crucial in managing intraprocedural complications during neurovascular embolization.
- It potentially reduces the need for emergency catheter upsizing and associated risks.