An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
A novel pedicle screw design to maximize screw-bone interface strength using finite element analysis and design of
Arvind Kumar Prajapati1, Parimanathukovilakom Ramavarma Harikrishna Varma1, Gurunathan Saravana Kumar2
1Bio-Medical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Thiruvananthapuram, India.
Study Design:
Basic study.
Purpose:
This study aimed to utilize finite element (FE) analysis and design of experiment (DoE) techniques to propose and optimize a novel pedicle screw design and compare its pull-out force with that of a control device.
Overview Of Literature:
Pedicle screw-based fixation is the gold-standard treatment for spine diseases, particularly in fusion procedures. However, pedicle screw loosening and breakage still occur in osteoporotic and non-osteoporotic patients. This research investigates screw design modifications to enhance screw-bone interface strength and reduce the likelihood of loosening.
Methods:
We conceptualized a novel pedicle screw considering vertebral bone morphology and strength differences. A validated FE model was developed and used in conjunction with DoE to determine the screw՚s optimum geometrical parameters. The FE model was validated through simulation and laboratory experiments using the control device. The optimized thread profiles for cortical bone and cancellous bone were determined, with pull-out force as the primary factor for screw design evaluation.
Results:
FE analysis results for the control device closely matched experimental results, with less than 5% difference. The chosen unique pitch/depth ratio showed maximum pull-out force for cortical bone, while DoE enabled the optimization of design parameters for cancellous bone. The optimized pedicle screw exhibited a 15% increase in pull-out force compared to the control device.
Conclusions:
The study proposes a novel pedicle screw design with better pull-out strength than the control device. Combining FE analysis with DoE is an effective approach for screw design optimization, reducing the need for extensive prototyping tests. A two-variable analysis suffices for optimizing cortical bone design parameters, while a multi-variable analysis is more effective for optimizing cancellous bone design parameters.
More Related Videos
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving
To evaluate the...
Design of Prismatic Beams for Bending
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...