Interobserver Reproducibility of Ultrasound Attenuation Imaging Technology in Liver Fat Quantification
Nurullah Dag1, Gulnur Igci1, Fatma Hilal Yagin2
1Faculty of Medicine, Department of Radiology, Inonu University, Malatya, Türkiye.
Purpose:
This study aims to investigate the interobserver variability in the quantitative assessment of liver fat content using ultrasound attenuation imaging technology (USAT).
Methods:
This prospective, single-center study included 96 adult patients who were either diagnosed with or suspected of having metabolic dysfunction-associated steatotic liver disease. Independent observers, blinded to each other's assessments, evaluated hepatic steatosis visually and through USAT measurements. Separate measurements were taken at five intercostal and subcostal sites, and the median values of these measurements were recorded. The correlation between USAT measurements and visual steatosis grades was examined using Spearman's correlation test. Intraclass correlation coefficient (ICC) and Bland-Altman analysis were used to evaluate the interobserver variability of USAT measurements.
Results:
Interobserver agreement for USAT measurements was excellent for the intercostal examination and good for the subcostal examination (p < 0.001). Body mass index did not significantly affect the level of interobserver agreement. Interobserver variability in Bland-Altman plots of USAT measurements was within the 95% limits of agreement. USAT measurements correlated very strongly with the visual degree of hepatic steatosis, both intercostal and subcostal (p < 0.001). USAT measurements were also significantly different between different visual degrees of hepatic steatosis (p < 0.001).
Conclusion:
In the assessment of hepatic steatosis, USAT measurements obtained from the intercostal space showed excellent agreement in terms of interobserver reproducibility.
More Related Videos
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...