Anesthetic Gas Scavenging System for Gas Evacuation in the ICU
Lucie Collet1, Mona Assefi1, Jean-Michel Constantin1
1The authors are affiliated with Sorbonne Université, GRC 29, DMU DREAM, AP-HP, Hôpital Pitié-Salpetrière, Département d'Anesthésie-Réanimation, Paris, France.
Abstract:
Background: Inhaled sedation is increasing in ICUs, with active carbon filters (ACFs) commonly used for evacuating halogenated gases. However, the potential benefits of a waste anesthetic gas system (WAGS) similar to the ones used in operating rooms should be explored. To limit the suction over the flow sensor where the WAGS is connected on ICU ventilators, an anesthetic gas receiving system (AGRS) is required, constituting with the WAGS an active gas receiving and scavenging system (AGRSS). Ensuring that this whole device does not compromise the flow sensor reliability is crucial. The aim of this study was to compare various gas evacuation devices and assess the reliability of AGRSS on ICU ventilators. Methods: In this experimental study, pressures and flows were recorded during the ventilation of a test lung using various ventilator settings and gas evacuation methods: no device (reference condition), ACF, the WAGS alone, AGRSS (WAGS and AGRS together), and the expiratory valve connected to the medical vacuum system with the AGRS in between. Visual comparisons of the pressure and flow curves followed by a statistical analysis comparing median pressures and flows of each device to the reference were performed. Results: The test lung model demonstrated consistent comparability in pressures and flows among all devices, except for the WAGS alone, which exhibited discordance through significant overestimation or underestimation. Conclusions: These findings indicate that using a WAGS with the AGRS system appeared to be reliable for managing gas evacuation in ICUs without compromising pressure or flow delivery. The data from this experimental trial should be confirmed with clinical studies involving human subjects. Given the increasing use of inhaled sedation in ICUs, these results support the daily application of the WAGS with the AGRS for gas evacuation, similar to its established use in anesthesiology.
More Related Videos
04:56In Vitro Method to Control Concentrations of Halogenated Gases in Cultured Alveolar Epithelial Cells
Published on: October 23, 2018
14:52Recording Brain Electromagnetic Activity During the Administration of the Gaseous Anesthetic Agents Xenon and Nitrous Oxide in Healthy Volunteers
Published on: January 13, 2018
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Inhalational Anesthetics: Overview
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...