Development and Validation of a Customized Amplex UltraRed Assay for Sensitive Hydrogen Peroxide Detection in
Alexandra H Heussner1, Marlene Hermann2, Melanie Zerulla-Wernitz2
1Analytical Science Laboratory, Development Service, Vetter Pharma-Fertigung GmbH & Co. KG, Schuetzenstrasse 87, 88212 Ravensburg, Germany alexandra.heussner@vetter-pharma.com.
Abstract:
For clean-room technologies such as isolators and restricted access barrier systems (RABS), decontamination using hydrogen peroxide (H2O2) is increasingly attractive to fulfill regulatory requirements. Several approaches are currently used, ranging from manual wipe disinfection to vapor phase hydrogen peroxide (VPHP) or automated nebulization sanitization. Although the residual airborne H2O2 concentration can be easily monitored, detection of trace H2O2 residues in filled products is rather challenging. To simulate the filling process in a specific clean room, technical runs with water for injection (WfI) are popular. Thus, the ability to detect traces of H2O2 in water is an important prerequisite to ensure a safe and reliable use of H2O2 for isolator or clean room decontamination. The objective of this study was to provide a validated quantitative, fluorometric Amplex UltraRed assay, which satisfies the analytical target profile of quantifying H2O2 in WfI at low nanomolar to low micromolar concentrations (ppb range) with high accuracy and high precision. The Amplex UltraRed technology provides a solid basis for this purpose; however, no commercial assay kit that fulfills these requirements is available. Therefore, a customized Amplex UltraRed assay was developed, optimized, and validated. This approach resulted in an assay that is capable of quantifying H2O2 in WfI selectively, sensitively, accurately, precisely, and robustly. This assay is used in process development and qualification approaches using WfI in H2O2-decontaminated clean rooms and isolators.
More Related Videos
05:17Fast and Specific Assessment of the Halogenating Peroxidase Activity in Leukocyte-enriched Blood Samples
Published on: July 28, 2016
08:57Simultaneous Measurement of Superoxide/Hydrogen Peroxide and NADH Production by Flavin-containing Mitochondrial Dehydrogenases
Published on: February 24, 2018