Background:
Fecal immunochemical tests are commonly performed for colorectal cancer screening. Instant fecal occult blood measurement in toilet bowel movements would improve convenience. Hyperspectral imaging (HSI) enables the nondestructive evaluation of materials that are difficult to assess visually. This study aimed to determine whether HSI could be used to identify fecal occult blood on stool surfaces.
Methods:
The study included 100 patients who underwent colonoscopy, divided into groups A and B (50 patients, each) for creating a discriminant algorithm and validating the accuracy of the algorithm, respectively. In group A, 100 areas were randomly selected from the stool surface, and the fecal occult blood quantitative values were measured and photographed using a hyperspectral camera (cutoff: > 400 ng/mL). A discriminant algorithm image was created to extract spectral feature differences obtained from HSI via machine learning. In group B, 250 random areas were evaluated and compared to fecal occult blood quantitative values, measuring sensitivity, specificity, accuracy, positive predictive value (PPV), and negative predictive value (NPV).
Results:
Groups A and B comprised 28 and 26 patients with cancer, respectively. Cancer detection sensitivity at the 400 ng/mL cutoff was 67.9% and 42.3% in groups A and B, respectively. The discriminant algorithm image exhibited high accuracy in group A (sensitivity; 77.1%, specificity; 96.9%, accuracy; 90.0%, PPV; 93.1%, NPV; 88.7%). In group B, the sensitivity, specificity, accuracy, PPV, and NPV were 83.3, 92.9, 90.8, 76.3, and 95.3%, respectively.
Conclusion:
HSI can effectively discriminate high quantitative fecal occult blood, highlighting its potential for improved colorectal cancer screening.