Automated deep learning-based bone mineral density assessment for opportunistic osteoporosis screening using various
Heejun Park1, Woo Young Kang2, Ok Hee Woo1
1Department of Radiology, Guro Hospital, Korea University Medical Center, Seoul, Republic of Korea.
Abstract:
This retrospective study examined the diagnostic efficacy of automated deep learning-based bone mineral density (DL-BMD) measurements for osteoporosis screening using 422 CT datasets from four vendors in two medical centers, encompassing 159 chest, 156 abdominal, and 107 lumbar spine datasets. DL-BMD values on L1 and L2 vertebral bodies were compared with manual BMD (m-BMD) measurements using Pearson's correlation and intraclass correlation coefficients. Strong agreement was found between m-BMD and DL-BMD in total CT scans (r = 0.953, p < 0.001). The diagnostic performance of DL-BMD was assessed using receiver operating characteristic analysis for osteoporosis and low BMD by dual-energy x-ray absorptiometry (DXA) and m-BMD. Compared to DXA, DL-BMD demonstrated an AUC of 0.790 (95% CI 0.733-0.839) for low BMD and 0.769 (95% CI 0.710-0.820) for osteoporosis, with sensitivity, specificity, and accuracy of 80.8% (95% CI 74.2-86.3%), 56.3% (95% CI 43.4-68.6%), and 74.3% (95% CI 68.3-79.7%) for low BMD and 65.4% (95% CI 50.9-78.0%), 70.9% (95% CI 63.8-77.3%), and 69.7% (95% CI 63.5-75.4%) for osteoporosis, respectively. Compared to m-BMD, DL-BMD showed an AUC of 0.983 (95% CI 0.973-0.993) for low BMD and 0.972 (95% CI 0.958-0.987) for osteoporosis, with sensitivity, specificity, and accuracy of 97.3% (95% CI 94.5-98.9%), 85.2% (95% CI 78.8-90.3%), and 92.7% (95% CI 89.7-95.0%) for low BMD and 94.4% (95% CI 88.3-97.9%), 89.5% (95% CI 85.6-92.7%), and 90.8% (95% CI 87.6-93.4%) for osteoporosis, respectively. The DL-based method can provide accurate and reliable BMD assessments across diverse CT protocols and scanners.
More Related Videos
06:59Author Spotlight: An Economic and Efficient Method for Quantitative Evaluation of Bone Microarchitecture in a Murine Osteoporosis Model
Published on: September 8, 2023
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025