Metal artifact reduction method based on single spectral CT (MARSS)
Zijing Zhu1, Haichuan Zhou2, Huitao Zhang1
1The school of Mathematical Sciences, Capital Normal University, Beijing, China.
Background:
For patients with metal implants, computed tomography (CT) imaging results suffer from metal artifacts, which seriously affect image evaluation and even lead to misdiagnosis. Because spectral CT technology has the advantage of quantitative imaging, basis material decomposition, and so on, the current metal artifact reduction methods are utilizing spectral information to reduce metal artifacts with good results. However, they usually require projection data from multiple spectra or energy-windows, which is difficult to realize in conventional CT.
Purpose:
To satisfy the status quo, the aim of this work is to propose a metal artifact reduction (MAR) method based on single spectral CT (MARSS). By incorporating prior information, the average density of some base materials, and a constrained image reconstruction model is established. It forces the solution spaces of the materials to be discrete and finite, making the model easier to solve.
Methods:
The MARSS method uses the idea of discrete tomography to establish a constrained reconstruction model. By incorporating priori knowledge, the constraint forces the solution spaces of some materials to be discrete, which can effectively downsize the solution space and reduce the ill-posedness of this problem. Then, an iteration algorithm is developed to solve this model. This algorithm iterates alternately between reconstruction and discretization. It ensures that the solution spaces are discrete while the polychromatic projection of the reconstructed image converges to that of the scanned object.
Results:
The MRASS method significantly reduces artifacts and restores structures near metal to a large extent. Unlike the comparison MAR methods, it effectively prevents the introduction of new artifacts and distortion of the structure.
Conclusions:
The MARSS method can achieve MAR based on single spectral CT. Subjective and quantitative evaluation of the results show that the method significantly improves image quality compared to competing methods.
More Related Videos
07:24Quantitative Atomic-Site Analysis of Functional Dopants/Point Defects in Crystalline Materials by Electron-Channeling-Enhanced Microanalysis
Published on: May 10, 2021
10:06Characterization Of Multi-layered Fish Scales Atractosteus spatula Using Nanoindentation, X-ray CT, FTIR, and SEM
Published on: July 10, 2014
Related Concept Videos
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...