Severe intraocular lens tilt after the Yamane technique
Neil Kelkar1, Kevin Eid, Austin S Nakatsuka
1From the Intermountain Ocular Research Center, Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah (Kelkar, Eid, Mamalis, Werner); Alan S. Crandall Center for Glaucoma Innovation, Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah (Nakatsuka, Johnson).
This study examined explanted CT Lucia 602 intraocular lenses, finding deformed or broken haptics and a rotisserie effect in some Yamane-technique implants, particularly from later batches.
Area of Science:
- Ophthalmology
- Medical Device Analysis
- Surgical Complications
Background:
- Intraocular lenses (IOLs) are crucial for vision restoration after cataract surgery.
- Implantation without adequate capsular support presents unique challenges.
- CT Lucia 602 IOLs have been used in complex cases, necessitating evaluation of their performance and failure modes.
Purpose of the Study:
- To evaluate explanted CT Lucia 602 intraocular lenses (IOLs) that were implanted without capsular support.
- To document the condition of the IOL haptics and assess for the 'rotisserie effect' after explantation.
- To correlate clinical findings with pathological examination of the explanted devices.
Main Methods:
- A case series involving 13 explanted CT Lucia 602 IOLs, primarily implanted using the Yamane technique.
- Clinical data review for cases with decentration, dislocation, or optic tilt.
- Gross and microscopic examination of explanted lenses to assess haptic integrity and rotation, under both dry and hydrated conditions.
Main Results:
- A significant proportion of haptics were found to be deformed (15/26) or broken (6/26).
- Four haptics from Yamane-technique lenses explanted after November 2022 exhibited free rotation after hydration, indicating the rotisserie effect.
- Lenses explanted later showed severe optic tilt, suggesting a correlation with haptic issues and the rotisserie effect.
Conclusions:
- The rotisserie effect in CT Lucia 602 IOLs implanted via the Yamane technique is a documented phenomenon.
- Findings suggest that specific manufacturing batches may be associated with increased risk of haptic failure and rotation.
- This study highlights potential complications with CT Lucia 602 IOLs in cases lacking capsular support, particularly concerning haptic integrity and rotational stability.