Tooth segmentation by low-dose CBCT for orthodontic treatment planning : Explorative ex vivo validation
Maurice Ruetters1, Holger Gehrig2, Sinclair Awounvo3
1Heidelberg University, Department of Operative Dentistry, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany. Maurice.ruetters@med.uni-heidelberg.de.
Summary
Low-dose cone-beam computed tomography (LD-CBCT) offers precise and reproducible measurements for tooth length, comparable to standard high-dose CBCT (HD-CBCT). Further research is needed to assess the clinical relevance of volume differences observed.
Area of Science:
- Dentistry
- Radiology
- Medical Imaging
Background:
- Three-dimensional imaging is crucial in orthodontics, but standard high-dose cone-beam computed tomography (HD-CBCT) poses radiation risks.
- Advancements have led to low-dose CBCT (LD-CBCT) devices, aiming to reduce patient exposure.
Purpose of the Study:
- To evaluate the precision and reproducibility of LD-CBCT compared to HD-CBCT for dental imaging.
- To assess if LD-CBCT is as accurate as HD-CBCT in segmenting tooth roots and crowns and measuring tooth length.
Main Methods:
- HD-CBCT and LD-CBCT scans were acquired from four human cadaveric heads.
- Thirty single-rooted teeth were segmented twice by a single investigator, and tooth length was measured.
- Lin's concordance correlation coefficient (CCC) and intraclass correlation coefficient (ICC) were used to assess agreement and reliability.
Main Results:
- Excellent agreement was found between LD-CBCT and HD-CBCT for tooth length measurements (CCC = 0.991).
- Intrarater reliability for tooth length measurements was excellent for both modalities (ICC > 0.997).
- While volume measurements differed significantly (p < 0.001), agreement between modalities was excellent (CCC = 0.975) with high intrarater reliability (ICC > 0.997).
Conclusions:
- LD-CBCT is a valid alternative to HD-CBCT for measuring tooth length in orthodontic applications.
- The clinical significance of volume measurement differences between LD-CBCT and HD-CBCT requires further investigation through in vivo studies.
Keywords:
Low-dose cone-beam computed tomographyRadiation exposureThree-dimensional imagingTooth lengthTooth volume