Large improvement in RF magnetic fields and imaging SNR with whole-head high-permittivity slurry helmet for
Soo Han Soon1,2, Xin Li1, Matt Waks1
1Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA.
Purpose:
To optimize the design and demonstrate the integration of a helmet-shaped container filled with a high-permittivity material (HPM) slurry with RF head coil arrays to improve RF coil sensitivity and SNR for human-brain proton MRI.
Methods:
RF reception magnetic fields ( ) of a 32-channel receive-only coil array with various geometries and permittivity values of HPM slurry helmet are calculated with electromagnetic simulation at 7 T. A 16-channel transmit-only coil array, a 32-channel receive-only coil array, and a 2-piece HPM slurry helmet were constructed and assembled. RF transmission magnetic field ( ), , and MRI SNR maps from the entire human brain were measured and compared.
Results:
Simulations showed that averaged improvement with the HPM slurry helmet increased from 57% to 87% as the relative permittivity (εr) of HPM slurry increased from 110 to 210. In vivo experiments showed that the average improvement over the human brain was 14.5% with the two-piece HPM slurry (εr ≈ 170) helmet, and the average and SNR were improved 63% and 34%, respectively, because the MRI noise level was increased by the lossy HPM.
Conclusion:
The RF coil sensitivity and MRI SNR were largely improved with the two-piece HPM slurry helmet demonstrated by both electromagnetic simulations and in vivo human head experiments at 7 T. The findings demonstrate that incorporating an easily producible HPM slurry helmet into the RF coil array significantly enhances human-brain MRI SNR homogeneity and quality at ultrahigh field. Greater SNR improvement is anticipated using the less lossy HPM and optimal design.