Abstract:
Efficient sampling materials are essential for assessing nicotine levels in vape shops and other settings where nicotine exposures may exist. Two different treatments of Whatman glass fiber type A (GF/A) filters (sodium bisulfate treated and citric acid treated) were evaluated for nicotine capture, desorption efficiency, and holding capacity using Gas Chromatography-Mass Spectrometry (GC-MS). The Filters were treated with 0.8 mL of 0.1 M sodium bisulfate or citric acid solution and oven-dried (80 °C) for 30 min. Nicotine was desorbed off the filters using a modified analytical method. The average nicotine desorption efficiency for sodium bisulfate-treated GF/A filters (98.4%) was significantly higher than that of citric acid-treated GF/A filters (60.9%) over a range of 1-100 µg nicotine. Sodium bisulfate-treated and citric acid-treated GF/A filters experienced a 10% nicotine breakthrough after being dosed with about 550 and 2,750 µg of nicotine, respectively compared to 75 µg for untreated GF/A filters. Citric acid-treated GF/A filters had a much greater nicotine-holding capacity, but nicotine desorption from citric acid-treated GF/A filters was below the recommended criteria. Therefore, we recommend that sodium bisulfate-treated GF/A filters are employed for sample of nicotine with the GC-MS method.
More Related Videos
09:30A Microcontroller Operated Device for the Generation of Liquid Extracts from Conventional Cigarette Smoke and Electronic Cigarette Aerosol
Published on: January 18, 2018
08:39Generation of Electronic Cigarette Aerosol by a Third-Generation Machine-Vaping Device: Application to Toxicological Studies
Published on: August 25, 2018
Related Concept Videos
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...