Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices
Hugo Lanz1,2, Katharina Strauß1,2, Julia Höpler3
1Department of Medicine I, LMU University Hospital, 81377 Munich, Germany.
Background:
MRI (magnetic resonance imaging) represents the diagnostic image modality of choice in several conditions. With an increasing number of patients requiring MRI for diagnostic purposes, the issue of safety in patients with cardiac implantable electronic devices (CIED) undergoing this imaging modality will play an ever more important role. The purpose of this study was to assess the safety and device function following MRI in an unrestricted real-world cohort of patients with a wide array of cardiac devices.
Methods:
We conducted a retrospective single-center study including 1010 MRI studies conducted in adult patients (≥18 years) with an implanted CIED treated in the University Hospital of Munich (LMU) between July 2012 and March 2024. Patients with non-MR conditionally labeled leads, abandoned or epicardial leads, as well as lead fragments, were included for analysis.
Results:
Across a total of 1010 MRIs (920 total MR-conditional device generators) performed in patients with an implanted CIED, there were no deaths, reports of discomfort, palpitations, heating, or ventricular arrythmias in the 24 h following MRI. Only 2/1010 MRIs were followed by a reported atrial arrhythmia within 24 h, both in patients with an MR-conditional pacemaker (PM) device without an abandoned lead. No significant changes in device function following MRI from baseline were observed across all included CIEDs. Lastly, no instances of severe malfunction, such as generator failure, loss of capture, electrical reset, or inappropriate inhibition of pacing, were found in post-MRI interrogation reports across all MRI studies.
Conclusions:
Based on the analysis of 1010 MRIs undergone by patients with CIEDs, following standardized device interrogation, manufacturer-advised device programming, monitoring of vital function, and manufacturer-advised reprogramming, MRI can be performed safely and without adverse events or changes in device function.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Holter Monitor: 24-Hour Monitoring