Related Experiment Video
Updated: May 6, 2026

09:36
Measurement of Spatial Stability in Precision Grip
Published on: June 4, 2020
3.1K
Isometric Force-Time Curve Assessment: Accuracy, Precision, and Repeatability of a Mobile Application and Portable
Dario Santos1,2, Alfredo Bravo-Sánchez3,4, Leonardo Alexandre Peyré-Tartaruga5,6
1Unidad Académica de Fisioterapia, Facultad de Medicina, Universidad de la República, Montevideo 11600, Uruguay.
Sports (Basel, Switzerland)
|October 25, 2024
Summary
A new portable device, DINABANG, accurately measures knee extensor strength. It shows excellent repeatability and reliability compared to the BIODEX isokinetic dynamometer for both maximal voluntary isometric contractions and paired voluntary isometric contractions.
Area of Science:
- Biomechanics and Exercise Physiology
- Sports Science and Rehabilitation Technology
Background:
- Strength assessment is crucial in sports performance, physical rehabilitation, and general health.
- Accurate and reliable measurement tools are essential for evaluating muscle function.
Purpose of the Study:
- To compare the accuracy and reliability of a portable DINABANG device against a standard BIODEX isokinetic dynamometer for knee extensor strength assessment.
- To evaluate both maximal voluntary isometric contractions (MVICs) and paired voluntary isometric contractions (VICs) using both devices.
Main Methods:
- Nineteen healthy volunteers participated in the study.
- Maximal voluntary isometric contractions (MVICs) and paired voluntary isometric contractions (VICs) of knee extensors were measured using both BIODEX and DINABANG.
- Repeatability, accuracy (bias), and precision (limits of agreement) were statistically analyzed using ICC and Bland-Altman plots.
Main Results:
- Excellent repeatability (ICC: 1.00) and trivial effect were found for MVICs between BIODEX and DINABANG, with high accuracy (bias: 0.12 Nm) and consistent precision.
- Excellent repeatability (ICC: 1.00) and trivial effect were also observed for VICs, with high accuracy (bias: 0.24 Nm) and consistent precision.
- The portable DINABANG device demonstrated comparable performance to the BIODEX system.
Conclusions:
- The DINABANG device is accurate, precise, and reliable for measuring knee extensor torque.
- This portable device offers a viable alternative for strength assessment in various settings, including sports performance and rehabilitation.