Different ECLS Pump Configurations for Temporary Right Ventricular Assist Device in LVAD Patients: A Retrospective
Dragan Opacic1,2,3, Christian Klüß1, Darko Radakovic1,2,3
1Clinic for Thoracic and Cardiovascular Surgery, Heart and Diabetes Centre North Rhine Westphalia, 32545 Bad Oeynhausen, Germany.
Life (Basel, Switzerland)
|October 26, 2024
Summary
Temporary right ventricular assist devices (RVADs) can cause bleeding complications after left ventricular assist device (LVAD) implantation. The Deltastream™ system showed higher bleeding rates and prolonged ICU stays compared to Centrimag™ and Cardiohelp™.
Area of Science:
- Cardiovascular Surgery
- Mechanical Circulatory Support
- Critical Care Medicine
Background:
- Acute right ventricular failure is a significant complication following left ventricular assist device (LVAD) implantation.
- Temporary paracorporeal right ventricular assist devices (RVADs) are commonly used for management.
- Different extracorporeal life support (ECLS) systems may impact patient outcomes.
Purpose of the Study:
- To compare the outcomes of three temporary RVAD systems: Centrimag™, Cardiohelp™, and Deltastream™.
- To evaluate mortality, bleeding complications, and intensive care unit (ICU) stay duration.
- To identify potential differences in system performance and patient impact.
Main Methods:
- A monocentric, retrospective case-control study included 245 patients who received LVAD with paracorporeal RVAD between 2009 and 2020.
- Patients were divided into three groups based on the RVAD system used: Centrimag™ (A), Cardiohelp™ (B), and Deltastream™ (C).
- Outcomes assessed included in-hospital mortality, reoperation rates due to bleeding, and ICU length of stay.
Main Results:
- While preoperative INTERMACS scores were similar between Centrimag™ and Deltastream™ groups, Cardiohelp™ patients had worse scores.
- In-hospital death rates were 41.8% (A), 32.6% (B), and 54.7% (C).
- Reoperation rates due to bleeding were significantly higher in the Deltastream™ group (47.2%) compared to Centrimag™ (21.9%) and Cardiohelp™ (17.4%), with increased thrombocyte consumption and prolonged ICU stays.
Conclusions:
- Temporary RVADs are associated with bleeding complications that affect patient outcomes.
- The Deltastream™ system demonstrated a significantly higher incidence of bleeding complications, potentially linked to high pump speeds and reduced thrombocyte counts.
- Further prospective studies are recommended to validate these findings due to the retrospective nature and patient complexity.
Keywords:
acute right ventricular failurebleedingextracorporeal life support pump configurationsleft ventricular assist deviceparacorporeal right ventricular assist device