Evaluation of extra-corporeal membrane oxygenator cannulae in pulsatile and non-pulsatile pediatric mock circuits
Lorenzo Ferrari1, Maris Bartkevics2, Hansjörg Jenni2
1ARTORG Center for Biomedical Engineering Research, University of Bern, Bern, Switzerland.
Background:
This study evaluated the hemodynamic performance of arterial and venous cannulae in a compliant pediatric extracorporeal membrane oxygenation (ECMO) mock circuit in pulsatile and non-pulsatile flow conditions.
Methods:
The ECMO setup consisted of an oxygenator, diagonal pump, and standardized-length arterial/venous tubing with pressure transducers. A validated left-heart mock loop was adapted to simulate pediatric conditions. The pulsatile flow was driven by a computer-controlled piston pump set at 120 bpm. A roller pump was used for non-pulsatile conditions. The circuit was primed with 40% glycerol-based solution. The cardiac output was set to 1 L/min and the aortic pressure to 40-50 mmHg. Four arterial cannulae (8Fr, 10Fr, 12Fr, 14Fr) and five venous cannulae (12Fr, 14Fr, 16Fr, 18Fr, 20Fr) (Medtronic, Inc., Minneapolis, MN, USA) were tested at increasing flow rate in 12 combinations.
Results:
The pulsatile condition required lower ECMO pump speeds for all cannulae combinations at a given flow rate, inducing a significantly smaller increase of flow in the mock loop. Under non-pulsatile conditions, the aortic and arterial pressures in the cannulae were higher (p < 0.01) while no significant differences in pressure drop and pressure-flow characteristics (M-number) were observed. The total hemodynamic energy was higher in case of non-pulsatile flow (p < 0.01).
Conclusion:
Under non-pulsatile conditions, the system was characterized by overall higher pressures, resulting in higher support to the patient. The consequent increase of potential energy compensates for increases of kinetic energy, leading to a higher total hemodynamic energy. Pressure gradients and M number are independent of the testing conditions. Pulsatile testing conditions led to more physiological testing conditions, and it is recommended for ECMO testing.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)