Pediatric contrast-enhanced chest CT on a photon-counting detector CT: radiation dose and image quality compared to
Alexander M El-Ali1,2, Naomi Strubel3,4, Lynne Pinkney3,4
1New York University Langone Medical Center, New York, USA. alexander.el-ali@nyulangone.org.
Background:
Photon counting detector (PCD) CT benefits from reduced noise compared with conventional energy-integrating detector (EID) CT, which should translate to improved image quality and reduced radiation exposure for pediatric patients undergoing chest CT with IV contrast.
Objective:
To determine the differences in radiation exposure and image quality of PCD CT and EID CT in pediatric chest CT with intravenous (IV) contrast.
Materials And Methods:
In this institutional review board-approved retrospective observational study, 20 scan pairs (20 PCD CT; 20 EID CT) for children who underwent chest CT with IV contrast on both a PCD CT (Siemens NAEOTOM Alpha) and an EID CT (Siemens SOMATOM Definition Edge or Force) within 12 months were reviewed independently by three pediatric radiologists for three subjective quality features on 5-point Likert scales: overall quality, small structure delineation, and motion artifact. Objective measures of image quality (image noise, signal-to-noise ratio, and contrast-to-noise ratio) were assessed by a single radiologist in several locations in the chest through region of interest measurement of Hounsfield units (HU) and standard deviation. Patient-related and radiation exposure parameters were collected for each scan and summarized with median and interquartile range (IQR). The Wilcoxon rank-sum test was utilized to compare groups. A P < 0.05 indicated statistical significance. Inter-observer agreement of subjective image quality metrics was analyzed using weighted kappa.
Results:
Age (14.2 years vs 13.8 years, P= 0.15), height (P= 0.13), weight (P= 0.21), and BMI (P = 0.24) did not significantly differ between groups. There were 10 male and 3 female patients. Compared to EID CT, PCD CT showed lower radiation exposure parameters including volumetric CT dose index, 1.7 mGy (IQR 1.1-2.4 mGy) vs 3.8 mGy (IQR 2.0-4.7 mGy) (P< 0.01), and size-specific dose estimate, 2.6 mGy (IQR 1.8-3.1 mGy) vs 5.0 mGy (IQR 3.3-6.2 mGy) (P< 0.01). Objective image quality of lung parenchyma was improved on the PCD CT scanner, including image noise 119.5 HU (IQR 95.4-135.7 HU) vs 143.1 HU (IQR 125.4-169.8 HU) (P < 0.01), signal-to-noise ratio (SNR) -6.1 (IQR -8.4 to -4.8) vs -4.9 (IQR -5.6 to -3.8) (P= 0.01), and contrast-to-noise ratio -63.9 (-84.1 to -57.5) vs -60.5 (-76.3 to -52.5) (P = 0.01). Motion artifact was improved on the PCD CT scanner (P< 0.01). No significant differences in overall image quality or small structure delineation were identified (P= 0.06 and P= 0.31).
Conclusion:
PCD CT pediatric chest CT had significantly reduced radiation exposure, improved image quality, and reduced motion artifact compared with EID CT.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
X-ray Imaging
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET