Comparison of cardiac indices during pregnancy between noninvasive automated device and two-dimensional
Piya Chaemsaithong1, Man Yan Chung1, Qian Sun1
1Department of Obstetrics and Gynaecology, Prince of Wales Hospital, The Chinese University of Hong Kong, Shatin, Hong Kong.
Objective:
To determine the correlation, precision, mean percentage difference and agreement between cardiac indices (stroke volume [SV], cardiac output [CO], SV index [SVI], and cardiac index [CI]) measured by noninvasive cardiac output monitor (NICOM, Cheetah Medical, Boston, MA, USA) and 2-dimensional transthoracic echocardiography (2D-TTE) across gestations in Chinese pregnant women.
Methods:
This was a prospective longitudinal study performed in women with singleton pregnancy at 11-14+6 (n = 152), 19-24+6 (n = 152), 30-34+6 (n = 141), and 35-37+6 (n = 103). Cardiac indices, including CO and SV, were obtained by NICOM®, which uses thoracic bioreactance, and 2D-TTE. CI and SVI were calculated from CO and SV adjusted for body surface area. The measurements of cardiac indices obtained using NICOM were assessed relative to that of 2D-TTE by calculating correlation coefficient, bias, precision, mean percentage difference, and 95% limits of agreement, adjusted for repeated measurements.
Results:
Comparison of the SV and SVI measurements by the two approaches showed significant moderate correlation in the first trimester (r = 0.2-0.3; p = .01). Overall, the SV and SVI measurements obtained using NICOM relative to that obtained by 2D-TTE revealed a bias of -12.1 mL and -6.1 mL/m2 (95% confidence interval [CI]: -44.5 to 20.2 and -24.8 to 12.5), respectively. Comparison of the CO and CI measurements by the two approaches showed significant moderate correlation in the first trimester (r∼0.2; p = .01). Overall, the CO and CI measurements obtained using NICOM relative to that obtained by 2D-TTE revealed a bias of -0.50 L/min and -0.18 L/min/m2 (95% CI: 2.26-3.27 and -1.79 to 1.39), respectively. Mean percentage difference for all cardiac parameters in all three trimesters were more than 30%.
Conclusions:
In Chinese pregnant women, NICOM® has underestimated cardiac indices (SV, CO, SVI, and CI) compared to that measured by 2D-TTE. The mean percentage differences for all cardiac indices are more than 30%, which are higher than the clinically acceptable limit. Future research is needed to determine whether adjustment factors should be applied to the proprietary algorithms used by the NICOM® for the estimation of SV and CO in the Chinese pregnant population.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...