Variations in intraocular pressure during closed-system surgical procedures.

L C Moorhead, C D Armeniades

    Archives of Ophthalmology (Chicago, Ill. : 1960)
    |February 1, 1986
    PubMed
    Summary

    Intraocular pressure (IOP) can rapidly change during eye surgery, reaching up to 110 mm Hg. Monitoring infusion line pressure is unreliable for assessing true IOP, highlighting the need for direct IOP monitoring and control during ophthalmic procedures.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Wildfire Induces Changes in Receiving Waters: A Review With Considerations for Water Quality Management.

    Water resources research·2023
    Same author

    Hydrodynamic analysis of intraocular pressure changes during anterior chamber procedures.

    Journal of cataract and refractive surgery·1992
    Same author

    Injection of a silicone prepolymer system into the gallbladder of rabbits.

    Investigative radiology·1991
    Same author

    Evaluation of drug treatments for proliferative vitreoretinopathy using vitreous microtensiometry.

    Annals of ophthalmology·1991
    Same author

    Why we all should have living wills.

    Medical economics·1991
    Same author

    Computerization of medicine: a double-edged sword.

    Archives of ophthalmology (Chicago, Ill. : 1960)·1991

    Area of Science:

    • Ophthalmology
    • Surgical Technology
    • Biomedical Engineering

    Background:

    • Intraocular pressure (IOP) management is critical during ophthalmic surgery.
    • Current methods for monitoring IOP during surgery may be inadequate.

    Purpose of the Study:

    • To accurately measure true intraocular pressure (IOP) during anterior segment surgery and vitrectomy in rabbits.
    • To evaluate the reliability of infusion line pressure as an indicator of IOP during these procedures.

    Main Methods:

    • Miniature pressure sensors were inserted into the anterior chamber and vitreous of rabbit eyes.
    • Infusion fluid pressure was simultaneously measured using a sensor in the infusion line.
    • IOP and infusion line pressure were recorded during simulated surgical conditions.

    Related Experiment Videos

    Main Results:

    • Significant and rapid IOP fluctuations, up to 110 mm Hg, were observed during simulated surgical procedures.
    • Infusion line pressure changes did not correlate with actual IOP fluctuations.
    • Elevating the infusion bottle height had minimal effect on IOP when fluid egress was present.

    Conclusions:

    • Infusion line pressure is an unreliable indicator of true intraocular pressure during ophthalmic surgery.
    • Direct IOP measurement is essential for managing IOP during anterior segment surgery and vitrectomy.
    • Effective IOP control strategies are necessary to prevent potential ocular damage during surgery.