CT imaging using variable helical pitch scanning for lower extremity arterial disease: Reduced contrast medium dose,
Xiao-Shi Li1, Ji-Gang Geng2, Yin-Hu Zhu2
1Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, No. 227 YanTa West Road, Xi'an 710061, China; Department of Radiology, Xi'an Daxing Hospital, No. 353 LaoDong Road, Xi'an 710016, China.
Objectives:
This study aimed to explore the feasibility of reducing contrast medium (CM) volume, improving image quality and diagnostic accuracy using variable helical pitch (VHP) scanning for patients with lower extremity arterial disease (LEAD).
Materials And Methods:
Eighty patients who underwent lower extremity CT angiography (CTA) were prospectively enrolled and randomly assigned to either the VHP group (n = 40) or the conventional group (n = 40). Quantitative parameters and qualitative scores were compared between the two groups. Additionally, out of these patients, 72 arteries from 18 patients had DSA as the reference standard, and the diagnostic accuracy for the degree of vessel stenosis was assessed and compared.
Results:
In the VHP group, the contrast volume was significantly lower than in the conventional group (79.55 ± 11.87 mL vs. 89.63 ± 10.03 mL, p < 0.001), showing a reduction of 12.7 %. For all image quality characteristics, scores in VHP group were significantly superior to those in the conventional groups (all p < 0.05). Quantitative analysis revealed that images from the VHP group exhibited superior CT enhancement, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in the anterior tibial arteries (ATA) and dorsali pedis arteries (DPA) compared to the conventional group (all p < 0.001). Moreover, segment-based analysis showed the VHP group had significantly higher positive predictive value (PPV) and accuracy than the conventional group (PPV: 100 % vs. 76.19 %, p = 0.01; accuracy: 100 % vs. 84.38 %, p = 0.01, respectively).
Conclusions:
The implementation of the VHP protocol led to a 12.7 % decrease in contrast medium dosage compared to the conventional lower extremity CTA scanning protocol. Furthermore, it improved image quality and diagnostic accuracy, particularly for arteries below the knee.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...