Home-Based Monitor for Gait and Activity Analysis
Published on: August 8, 2019
Validity and reliability of running gait measurement with the ViMove2 system
Rachel Mason1, Gillian Barry1, Gary Hall1
1Department of Sport, Exercise and Rehabilitation, Northumbria University, Newcastle upon Tyne, United Kingdom.
The ViMove2 wearable system can measure running gait metrics like ground contact time and cadence, but accuracy is reduced at slower speeds and in female runners. Further research is needed for improved algorithm corrections.
Area of Science:
- Biomechanics
- Sports Science
- Wearable Technology
Background:
- Traditional running biomechanics analysis is lab-based, potentially not reflecting natural gait.
- Wearable technology offers potential for precise, real-world running gait monitoring.
Purpose of the Study:
- Evaluate the analytical validity and intra-session reliability of the ViMove2 wearable system for temporal running gait outcomes.
- Assess the accuracy of ViMove2 measurements against a 3D motion capture reference system.
Main Methods:
- Seventy-four healthy adults ran on a treadmill at various speeds (8-14 km/hr).
- Running gait was simultaneously measured using the ViMove2 wearable device and 3D motion capture (Vicon).
- Temporal gait outcomes analyzed included ground contact time (GCT) and cadence.
Main Results:
- ViMove2 showed expected changes in GCT and cadence with varying speeds, indicating face validity.
- ViMove2 tended to overestimate GCT and underestimate cadence, particularly at slower speeds.
- Agreement between ViMove2 and the reference was moderate to excellent for cadence across speeds, but varied for GCT, showing poor agreement at 8 km/hr and in females.
Conclusions:
- The ViMove2 system can measure temporal gait outcomes (GCT, cadence) in healthy adults.
- Accuracy and reliability are affected by slower running speeds and sex differences.
- Further research is required to develop speed- and sex-dependent corrections for ViMove2 algorithms.