Dosimetric study of bevel factors in IOERT with mobile linacs: Towards a unified code of practice
Rafael Ayala1, Rocío García2, Gema Ruiz2
1Servicio de Dosimetría y Radioprotección, Hospital General Universitario Gregorio Marañón, Dr. Esquerdo 46, 28007, Madrid, Spain; Nuclear Physics Group and IPARCOS, Department of Structure of Matter, Thermal Physics and Electronics, CEI Moncloa, Universidad Complutense de Madrid, Madrid, Spain.
Background:
Dosimetry in intraoperative electron radiotherapy (IOERT) poses distinct challenges, especially with inclined applicators deviating from international protocols. Ion recombination in ionization chambers, electron beam degradation due to scattering in cylindrical applicators, coupled with a lack of a well-defined beam quality surrogate, complicate output factor determination with ionization chambers. Synthetic diamond-based detectors, offer potential solutions; however, their suitability requires further exploration.
Purpose:
This study addresses output factor determination for beveled applicators. Objectives include assessing the suitability of PTW microDiamond detectors and determining correction factors for ionization chamber measurements based on energy variations at the depth of maximum dose (zmax) for beveled applicators. Experimental data are compared against results obtained from Monte Carlo simulations.
Methods:
We conducted measurements using both PTW microDiamond and IBA CC01 detectors. In addition to benchmarking bevel factors with penEasy, we employed Monte Carlo simulations to determine angular response correction factors for microDiamond detectors and to evaluate energy variations at zmax for beveled applicators.
Results:
The findings indicate that angular response correction factors are unnecessary for microDiamond detectors with beveled applicators. However, variations in mean energy at zmax potentially impact absorbed dose calculations with ionization chambers, particularly with the most inclined applicators.
Conclusions:
Based on our results, the study recommends using microDiamond detectors over cylindrical ionization chambers for output factor determination in IOERT with inclined applicators. Addressing energy variations at zmax is crucial to improve accuracy in ionization chamber measurements. These findings have implications for dosimetry protocols in IOERT, contributing to enhanced delivery in clinical practice.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiation: Applications
The average...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Standards of Care II
Absorption of Radiation
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...