Automated determination of urinary creatinine without sample dilution: theory and practice
Clinical Chemistry
|March 1, 1986
Summary
This study presents an automated method for determining urinary creatinine using the Jaffé reaction. The new method offers accurate and reliable results, serving as an effective replacement for existing automated analyzers.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Biochemistry
Background:
- The Jaffé reaction is a common method for creatinine determination.
- The reaction rate is influenced by sodium hydroxide concentration.
- Existing automated methods may require manual sample dilution.
Purpose of the Study:
- To develop and validate an automated method for direct urinary creatinine determination.
- To optimize the Jaffé reaction conditions for automated analysis.
- To compare the new method with established automated analyzers.
Main Methods:
- An automated kinetic assay was developed using the Jaffé reaction.
- Optimized conditions included 10 mmol/L picrate, 60 mmol/L sodium hydroxide, and a 1:41 sample to final volume ratio at 37°C.
- Measurements were taken at 500 or 510 nm between 30 and 90 seconds.
Main Results:
- The automated method provided results comparable to the AutoAnalyzer and aca.
- Run-to-run coefficients of variation (CVs) were 3.6% or less.
- The method demonstrated accuracy for creatinine concentrations up to 3000 mg/L.
Conclusions:
- The developed automated method allows for direct urinary creatinine determination without manual dilution.
- This method is a suitable and accurate replacement for the AutoAnalyzer and aca.
- The optimized Jaffé reaction offers a reliable approach for routine clinical analysis.