Safety and Feasibility of Interventional Hybrid Fluoroscopy and Nuclear Imaging in the Work-up Procedure of Hepatic
Martijn M A Dietze1, Marjolein B M Meddens1, Rob van Rooij1
1From the Department of Radiology and Nuclear Medicine, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX, PO Box 85500, 3508 GA Utrecht, the Netherlands.
Abstract:
Purpose To evaluate the safety and feasibility of a novel hybrid nuclear and fluoroscopy C-arm scanner to be used during the work-up procedure of hepatic radioembolization. Materials and Methods In this prospective first-in-human clinical study, 12 participants (median age, 67 years [range: 37-78 years]; nine [75%] male, three [25%] female) with liver tumors undergoing work-up for yttrium 90 radioembolization were included (ClinicalTrials.gov NCT06013774). Work-up angiography and technetium 99m-macroaggregated albumin injection were performed in an angiography suite equipped with a hybrid C-arm that could simultaneously perform fluoroscopy and planar nuclear imaging. Technetium 99m-macroaggregated albumin was injected under real-time hybrid imaging, followed by in-room SPECT imaging. Safety and feasibility were studied by assessing adverse events, technical performance, additional x-ray radiation dose, and questionnaires completed by radiologists and technologists. Results No adverse events were attributed to the hybrid C-arm scanner. The additional x-ray radiation dose was low (median, 19 Gy · cm2; minimum: 12 Gy · cm2; maximum: 21 Gy · cm2 for participants who completed all imaging steps). The interventional personnel considered use of the hybrid C-arm scanner safe and feasible, although the additional time spent in the intervention room was considered long (median, 64 minutes; minimum: 55 minutes; maximum: 77 minutes for participants who completed all imaging steps). Conclusion Use of the hybrid C-arm scanner during the work-up procedure of hepatic radioembolization was found to be safe and feasible in this first-in-human clinical study. Keywords: Angiography, Fluoroscopy, Interventional-Vascular, Radionuclide Studies, Radiosurgery, Gamma Knife, Cyberknife, SPECT, Instrumentation, Physics, Technical Aspects, Technology Assessment Supplemental material is available for this article. Published under a CC BY 4.0 license. Clinical trial registration no. NCT06013774.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
07:47Non-Invasive PET/MR Imaging in an Orthotopic Mouse Model of Hepatocellular Carcinoma
Published on: August 31, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET