Bias and precision of SPECT-based 177Lu activity-concentration estimation using a ring-configured solid-state versus
Anna Stenvall1, Irma Ceric Andelius1,2, Elias Nilsson1
1Radiation Physics, Department of Haematology, Oncology, and Radiation Physics, Skåne University Hospital, Lund, Sweden.
Background:
The aim was to compare bias and precision for 177Lu-SPECT activity-concentration estimation using a dual-headed Anger SPECT system and a ring-configured CZT SPECT system. This was investigated for imaging at 208 keV and 113 keV, respectively.
Methods:
Phantom experiments were performed on a GE Discovery 670 system with 5/8'' NaI(Tl) crystal (dual-headed Anger system) and a GE StarGuide (ring-configured CZT system). Six spheres (1.2 mL to 113 mL) in a NEMA PET body phantom were filled with 99mTc and 177Lu, separately. Mean relative errors and coefficients of variation (CV) in estimated sphere activity concentration were studied over six timeframes of 10 min each for the two systems. For 177Lu, similar acquisitions were also performed for an anthropomorphic phantom with two spheres (10 mL and 25 mL) in a liver with non-radioactive background and a sphere-to-background ratio of 15:1. Tomographic reconstruction was performed using OS-EM with 10 subsets with compensation for attenuation, scatter, and distance-dependent spatial resolution. For the Anger system, up to 40 iterations were used and for the ring-configured CZT system up to 30 iterations were used.
Results:
The two systems showed similar mean relative errors and CVs for 177Lu when using an energy window around 208 keV, while the ring-configured system demonstrated a lower bias for a similar CV compared to the Anger system for 99mTc and for 177Lu when using an energy window around 113 keV. However, total activity in the phantom tended to be overestimated in both systems for these cases.
Conclusions:
The ring-configured CZT system is a viable alternative to the dual-headed Anger system equipped with medium-energy collimators for 177Lu-SPECT and shows a potential advantage for activity-concentration estimation when operated at 113 keV. However, further consideration of the preservation of total activity is warranted.
More Related Videos
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
08:53Dependence of Laser-induced Breakdown Spectroscopy Results on Pulse Energies and Timing Parameters Using Soil Simulants
Published on: September 23, 2013
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Emission Spectroscopy: Overview
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...