Electrophysiologic characteristics and clinical correlation of right ventricular activation during left bundle branch
Shunmuga Sundaram Ponnusamy1, Vithiya Ganesan2, Vadivelu Ramalingam1
1Department of Cardiology, Velammal Medical College, Madurai, India.
Background:
Left bundle branch area pacing (LBBAP) results in a right bundle branch (RBB) delay pattern because of preexcitation of the left bundle. The mechanism of right ventricular (RV) activation during LBBAP is largely unknown.
Objective:
The aim of the study was to analyze the electrophysiologic characteristics of RV activation by mapping the RBB during LBBAP and its clinical correlation.
Methods:
Consecutive patients who underwent successful LBBAP were included. RBB block, RV paced rhythm, and suboptimal intracardiac electrograms were excluded. LBBAP was performed with continuous recording of His bundle (HB) and RBB electrograms. RV activation was classified into 3 types based on the intracardiac electrogram: type I, RBB mediated; type II, transseptal activation; and type III, fusion pattern.
Results:
Overall, 86 patients (94% left bundle branch pacing [LBBP]; 6% left ventricular septal pacing) were included. The mean age was 59.6 ± 12.8 years. Nonselective to selective capture transition was noted in 85% (n = 73). In patients with baseline normal QRS (n = 47), during selective LBBP (S-LBBP; n = 39), the most common pattern was type I (n = 34 [87%]), whereas during nonselective LBBP (NS-LBBP; n = 44), type III pattern (n = 40 [91%]) was common. In patients with left bundle branch block (n = 39), type III pattern was common during both S-LBBP and NS-LBBP. Type I pattern was noted only in patients with retrograde HB activation during S-LBBP. Left ventricular septal pacing showed type II activation in both groups. Patients without retrograde HB activation had higher left ventricular end-diastolic diameter, lower left ventricular ejection fraction, and prolonged HV interval compared with those with retrograde HB activation.
Conclusion:
Physiologic RBB-mediated (type I) activation of the right ventricle was the most common pattern observed during S-LBBP in patients with intact retrograde HB activation. Type III pattern was the most common pattern observed during NS-LBBP with fusion of multiple wavefronts from anterograde RBB activation, myocardial, and transverse interbundle connections.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and...