Back-end readout electronic design and initial results: a head-and-neck dedicated PET system based on CZT
Yuli Wang1, Ryan Herbst2, Shiva Abbaszadeh1
1Department of Electrical and Computer Engineering, University of California, Santa Cruz, Santa Cruz, CA, 95064.
Abstract:
Cadmium zinc telluride (CZT) radiation detectors are suitable for various applications, due to good energy performance at room temperature and simple pixilation to achieve high spatial resolution. Our group is developing a two-panel head-and-neck dedicated positron emission tomography (PET) system with CZT detectors. In this work, we present the back-end readout electronic design and the initial electronic noise results of our system. The back-end readout electronic incorporates RENA boards (a total 150 RENA boards for each panel), 30:1 fan-in boards connecting 30 RENA boards to the PicoZed 7010/7020 board (a total 5 fan-in boards for each panel). In each panel, 5:1 intermediate boards are used for biasing CZT detectors. The RENA board and the Picozed board are capable of data transmission of 50 Mbps and 6.6 Gbps, respectively. Electronic noise was also quantified using a square wave test pulse that provides charge injection into each channel of the RENA chip in the amount of 75fC/Volt. The pulse amplitude was chosen to generate approximately the same amount of charges as 511 keV photon would provide for each channels. The FWHM electronic noise at 511 keV was measured to be less than 1% (FWHM of 7.80 ± 1.47 ADC units or 4.89 ± 0.92 keV after conversion).
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET