[Long-term Reproducibility Evaluation of Fluoroscopy Dose Rate in Angiography Equipment]

Kazuya Mori1,2, Kouhei Makabe1, Ryou Sekiguchi1

  • 1Department of Radiological Technology, Saiseikai Kawaguchi General Hospital.

PubMed
Summary

This study monitored fluoroscopic dose rates in angiography systems over eight years. Results show stable X-ray output, with dose rate fluctuations within ±5%, ensuring reliable patient dose management.