What CPAP to use in the delivery room? Bench comparison of two methods to provide continuous positive airways
Viktoria Gruber1, Mark Brian Tracy2,3, Murray Kenneth Hinder3
1Department of Paediatrics and Adolescent Medicine, Division of Neonatology, Medical University of Graz, Graz, Steiermark, Austria.
Background:
Continuous positive airway pressure (CPAP) is a recommended first-line therapy for infants with respiratory distress at birth. Resuscitation devices incorporating CPAP delivery can have significantly different imposed resistances affecting airway pressure stability and work of breathing.
Aim:
To compare CPAP performance of two resuscitation devices (Neopuff T-piece resuscitator and rPAP) in a neonatal lung model simulating spontaneous breathing effort at birth.
Methods:
The parameters assessed were variation in delivered pressures (∆P), tidal volume (VT), inspiratory effort (model pressure respiratory muscle (PRM)) and work of breathing (WOB). Two data sequences were required with Neopuff and one with rPAP: (1) set PRM with changes in VT and (2) constant VT (preterm 6 mL, term 22 mL) with increased effort. Data were collected at CPAP settings of 5, 7 and 9 cmH2O using a 1 kg preterm (Compliance: 0.5 mL/cmH2O) and 3.5 kg term (1.0 mL/cmH2O) model.
Results:
2298 breaths were analysed (760 rPAP, 795 Neopuff constant VT, 743 Neopuff constant PRM). With CPAP at 9 cmH2O and set VT the mean ∆P (cmH2O) rPAP vs Neopuff 1.1 vs 5.6 (preterm) and 1.9 vs 13.4 (term), WOB (mJ) 4.6 vs 6.1 (preterm) and 35.3 vs 44.5 (term), and with set PRM mean VT (ml) decreased to 6.2 vs 5.2 (preterm) and 22.3 vs 17.5 (term) p<0.001. Similar results were found at pressures of 5 and 7 cmH2O.
Conclusion:
rPAP had smaller pressure swings than Neopuff at all CPAP levels and was thus more pressure stable. WOB was higher with Neopuff when VT was held constant. VT reduced with Neopuff when respiratory effort was constant.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...