Monte Carlo modeling of a commercial machine and experimental setup for FLASH-minibeam irradiations with electrons
Anthony Bonfrate1,2, Maria Grazia Ronga1,2, Annalisa Patriarca1
1Institut Curie, PSL Research University, Radiation Oncology Department, Proton Therapy Centre, Centre Universitaire, Orsay, France.
Background:
Ultra-high dose rate (UHDR/FLASH) irradiations, along with particle minibeam therapy (PMBT) are both emerging as promising alternatives to current radiotherapy techniques thanks to their improved healthy tissue sparing and similar tumor control.
Purpose:
Monte Carlo (MC) modeling of a commercial machine delivering 5-7 MeV electrons at UHDR. This model was used afterward to compare measurements against simulations for an experimental setup combining both FLASH and PMBT modalities.
Methods:
We modeled the main accelerator elements with TOPAS3.8/Geant4.10.07.p03, optimized the electron source parameters, and subsequently benchmarked this geometry against measurements. Minibeam experiments were performed by delivering 7 MeV electrons at UHDR on three different 65-mm thick brass collimators as manufactured for protons with a 400-µm slit width: single slit, 5 slits with a center-to-center (CTC) distance of 4 mm and 9 slits with CTC of 2 mm. Finally, complementary simulations were run by changing critical PMBT collimator parameters to assess their specific impact on peak-to-valley dose ratio (PVDR) as well as on the Bremsstrahlung photon contribution to the total dose.
Results:
Percentage depth dose (PDD) distributions and lateral dose profiles showed a good agreement between simulations and measurements, with a maximum discrepancy of less than 4%. With the PMBT collimators in place, discrepancies between simulated and measured dose profiles, lateral and in-depth in peaks and valleys, were within 3%. High PVDR between 5 and 26 were observed until 4 mm in the phantom. During the experiments, a mean dose rate of 167 Gy/s and an instantaneous dose rate of 1.2 × 105 Gy/s were obtained for the FLASH-minibeam setup. PMBT collimator parameters need to be optimized to maximize PVDR while limiting Bremsstrahlung photon contribution to the total dose.
Conclusions:
The validation of the MC model and the configuration of an electron FLASH-minibeam setup were successfully completed, paving the way for future radiobiological investigations.