Pulsation artifact reduction using compressed sensitivity encoding in Gd-EOB-DTPA contrast-enhanced magnetic
Masafumi Nakamura1,2, Yasuo Takatsu3,4, Mutsumi Yoshizawa5
1Department of Radiological Technology, Faculty of Health and Welfare, Tokushima Bunri University, 1314-1, Shido, Sanuki-City, Kagawa, 769-2193, Japan. nakamuraxx2@gmail.com.
Compressed sensitivity encoding (CS) significantly reduces cardiac pulsation artifacts in Gd-EOB-DTPA-enhanced MRI scans. This technique improves image quality, aiding in clearer diagnoses of liver conditions.
Area of Science:
- Medical Imaging
- Radiology
- Magnetic Resonance Imaging
Background:
- Cardiac pulsation artifacts in Gd-EOB-DTPA-enhanced MRI can impede accurate diagnosis in the left liver lobe.
- Improving image quality is crucial for effective diagnostic interpretation.
Purpose of the Study:
- To evaluate the efficacy of compressed sensitivity encoding (CS) in reducing cardiac pulsation artifacts.
- To enhance image quality in Gd-EOB-DTPA-enhanced three-dimensional (3D) T1-weighted turbo-field echo (3D-T1TFE) sequences.
Main Methods:
- Phantom studies using a metronome-synchronized cardiac phantom to acquire images with CS-eTHRIVE and S-eTHRIVE.
- Quantitative evaluation using standard deviation of difference images based on heart rate and TFE factor.
- Clinical evaluation of artifact reduction in hepatobiliary phase images through visual assessment.
Main Results:
- CS-eTHRIVE demonstrated significantly lower standard deviation values compared to S-eTHRIVE across all tested heart rates (P < 0.01).
- CS-eTHRIVE with a TFE factor of 35 showed the lowest standard deviation, significantly differing from other groups (P < 0.01).
- Clinical images acquired with CS-eTHRIVE received higher visual scores (3.72 ± 0.46) than those with S-eTHRIVE (2.72 ± 0.98, P < 0.01).
Conclusions:
- Compressed sensitivity encoding effectively reduces cardiac pulsation artifacts in Gd-EOB-DTPA-enhanced 3D-T1TFE MRI.
- This technique offers a significant improvement in image quality for liver imaging, aiding diagnostic accuracy.
More Related Videos
05:07Author Spotlight: Optimized Lung MRI Protocol with Computationally Efficient Reconstruction Methods
Published on: September 6, 2024
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
Related Concept Videos
Magnetic Resonance Imaging
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...